Spiral scan peripheral nerve stimulation.

K F King1, D J Schaefer.   

Abstract

Time-varying magnetic fields induce electric fields that can cause physiological stimulation. Stimulation has been empirically characterized as a function of dB/dt and duration based on experiments using trapezoidal and sinusoidal gradient waveforms with constant ramp time, amplitude, and direction. For two-dimensional (2D) spiral scans, the readout gradient waveforms are frequency- and amplitude-modulated sinusoids on two orthogonal axes in quadrature. The readout gradient waveform therefore rotates with amplitude and angular velocity that are generally not constant. It does not automatically follow that spiral stimulation thresholds can be predicted using available stimulation models. We scanned 18 normal volunteers with a 2D spiral scan and measured global thresholds for axial, sagittal, and coronal planes. We concluded that the stimulation model evaluated accurately predicts slew rate-limited spiral mean stimulation thresholds, if the effective ramp time is chosen to be the half-period at the end of the spiral readout.

Entities:  

Mesh:

Year:  2000        PMID: 10931576     DOI: 10.1002/1522-2586(200007)12:1<164::aid-jmri18>3.0.co;2-r

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  Increasing the oscillation frequency of strong magnetic fields above 101 kHz significantly raises peripheral nerve excitation thresholds.

Authors:  Irving N Weinberg; Pavel Y Stepanov; Stanley T Fricke; Roland Probst; Mario Urdaneta; Daniel Warnow; Howard Sanders; Steven C Glidden; Alan McMillan; Piotr M Starewicz; J Patrick Reilly
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

Review 2.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

3.  The effect of spiral trajectory correction on pseudo-continuous arterial spin labeling with high-performance gradients on a compact 3T scanner.

Authors:  Daehun Kang; Uten Yarach; Myung-Ho In; Erin M Gray; Joshua D Trzasko; Hang Joon Jo; Yunhong Shu; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2019-12-04       Impact factor: 4.668