Commissioning an image-guided localization system for radiotherapy.

M H Phillips1, K Singer, E Miller, K Stelzer.   

Abstract

PURPOSE: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use. METHODS AND MATERIALS: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system.
RESULTS: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system.
CONCLUSION: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault.

Entities:  

Mesh:

Year:  2000        PMID: 10924998     DOI: 10.1016/s0360-3016(00)00581-2

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Mitigating errors in external respiratory surrogate-based models of tumor position.

Authors:  Kathleen T Malinowski; Thomas J McAvoy; Rohini George; Sonja Dieterich; Warren D D'Souza
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-04-01       Impact factor: 7.038

Review 2.  Management of Radiation Proctitis.

Authors:  Lameese Tabaja; Shafik M Sidani
Journal:  Dig Dis Sci       Date:  2018-09       Impact factor: 3.199

Review 3.  Endoscopic and non-endoscopic approaches for the management of radiation-induced rectal bleeding.

Authors:  Joseph Paul Weiner; Andrew Thomas Wong; David Schwartz; Manuel Martinez; Ayse Aytaman; David Schreiber
Journal:  World J Gastroenterol       Date:  2016-08-21       Impact factor: 5.742

4.  Evaluation of patient setup uncertainty of optical guided frameless system for intracranial stereotactic radiosurgery.

Authors:  Jia-Zhu Wang; Roger Rice; Todd Pawlicki; Arno J Mundt; Ajay Sandhu; Joshua Lawson; Kevin T Murphy
Journal:  J Appl Clin Med Phys       Date:  2010-04-17       Impact factor: 2.102

5.  Design and study of ultrasound-based automatic patient movement monitoring device for quantifying the intrafraction motion during teletherapy treatment.

Authors:  S Senthilkumar; R Vinothraj
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

6.  Clinical Positioning Accuracy for Multisession Stereotactic Radiotherapy With the Gamma Knife Perfexion.

Authors:  Wade P Smith; Lori A Young; Mark H Phillips; Michael Cheung; Lia M Halasz; Jason K Rockhill
Journal:  Technol Cancer Res Treat       Date:  2017-05-17
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.