OBJECTIVE: The purpose of the study was to show that stereotactic breast biopsy can be performed effectively and accurately using add-on stereotactic equipment when it is performed with the patient in a decubitus or recumbent position.
MATERIALS AND METHODS: We retrospectively reviewed the results of 225 stereotactic core breast biopsies performed during a 42-month period. The biopsies were performed using a reclining mammography chair and add-on stereotactic equipment. Procedures were performed with the patient in the right or left lateral decubitus position or upright. Patients with benign biopsy results were followed up mammographically at 6 months initially, then yearly.
RESULTS: Of the 225 lesions biopsied stereotactically, 205 lesions (91%) were biopsied with the patient in the decubitus (right or left) position and 20 (9%) biopsied with the patient upright. Pathology results were classified as abnormal in 45 lesions (20%) and normal in 180 lesions (80%). All but one of the abnormal lesions were followed up with surgical excision. One hundred thirty-four of the 180 benign lesions have been followed up with at least one 6-month mammogram (follow-up range, 6-40 months), and none of the lesions has been subsequently proven malignant.
CONCLUSION: Stereotactic core biopsy can be performed effectively and accurately using standard add-on equipment when it is performed with the patient in the decubitus position. This approach obviates the use of dedicated prone stereotactic equipment.

Entities:  

Mesh:

Year:  2000        PMID: 10882246     DOI: 10.2214/ajr.175.1.1750059

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  6 in total

Review 1.  Stereotactic and ultrasound-guided breast biopsy.

Authors:  T H Helbich; W Matzek; M H Fuchsjäger
Journal:  Eur Radiol       Date:  2003-11-13       Impact factor: 5.315

2.  Lateral decubitus positioning stereotactic vacuum-assisted breast biopsy with true lateral mammography.

Authors:  Youn Joo Jung; Young Tae Bae; Jee Yeon Lee; Hyung-Il Seo; Jee Yeon Kim; Ki Seok Choo
Journal:  J Breast Cancer       Date:  2011-03-31       Impact factor: 3.588

3.  Stereotactic vacuum-assisted breast biopsy under lateral decubitus position.

Authors:  Sang Hyup Lee; Youn Joo Jung; Hyuk Jae Jung; Jee Yeon Kim; Ki Seok Choo; Kyung Jin Nam; Hyun Yul Kim
Journal:  Ann Surg Treat Res       Date:  2015-12-29       Impact factor: 1.859

4.  Breast Microcalcifications: Diagnostic Outcomes According to Image-Guided Biopsy Method.

Authors:  Sohi Bae; Jung Hyun Yoon; Hee Jung Moon; Min Jung Kim; Eun-Kyung Kim
Journal:  Korean J Radiol       Date:  2015-08-21       Impact factor: 3.500

5.  The clinical utility of a adding lateral approach to conventional vertical approach for prone stereotactic vacuum-assisted breast biopsy.

Authors:  Joo Hwa Myong; Bong Joo Kang; Soo Kyung Yoon; Sung Hun Kim; Yeong Yi An
Journal:  Korean J Radiol       Date:  2013-07-17       Impact factor: 3.500

6.  Stereotactic biopsy of the breast using a decubitus table: comparison of histologic underestimation rates between 11- and 8-gauge vacuum-assisted breast biopsy.

Authors:  Kyoung Eun Lee; Hak Hee Kim; Hee Jung Shin; Joo Hee Cha
Journal:  Springerplus       Date:  2013-10-22
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.