Literature DB >> 10794238

Early catastrophic failure of rotating hinge total knee prosthesis.

C J Wang1, H E Wang.   

Abstract

The mechanical failure of a prosthesis component is usually observed as a late complication of total knee arthroplasty (TKA). In knees with severe ligament instability and bone deficiency, either a true hinge prosthesis or a rotating hinge implant is commonly used. Failure of the polyethylene bearing bush in a hinge-type prosthesis is a complication that has not been reported to date. We report the cases of 2 rotating hinge TKA prostheses that dislocated as a result of mechanical failure of the prosthetic component within 5 months of initial implantation. Clinicians should be aware of this potential complication when selecting rotating hinge prostheses for certain patients.

Entities:  

Mesh:

Year:  2000        PMID: 10794238     DOI: 10.1016/s0883-5403(00)90877-5

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  14 in total

1.  Case report: Failure of rotating-hinge total knee prosthesis by disengagement of the hinge-post extension.

Authors:  Debdut Biswas; Bryan Haughom; Robert E Mayle; Craig J Della Valle
Journal:  Clin Orthop Relat Res       Date:  2012-12-11       Impact factor: 4.176

Review 2.  Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature.

Authors:  Joerg Friesenbichler; Ran Schwarzkopf; Patrick Sadoghi; Scott E Marwin; Mathias Glehr; Werner Maurer-Ertl; Andreas Leithner
Journal:  Int Orthop       Date:  2011-12-28       Impact factor: 3.075

3.  Primary total knee arthroplasty using rotating-hinge prosthesis in severely affected knees.

Authors:  Jae-Hyuk Yang; Jung-Ro Yoon; Chi-Hun Oh; Taik-Sun Kim
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-07-20       Impact factor: 4.342

4.  Hinged total knee arthroplasty in the presence of ligamentous deficiency.

Authors:  Daniel Hernández-Vaquero; Manuel A Sandoval-García
Journal:  Clin Orthop Relat Res       Date:  2010-05       Impact factor: 4.176

5.  Distal femoral replacement in nontumor cases with severe bone loss and instability.

Authors:  Keith R Berend; Adolph V Lombardi
Journal:  Clin Orthop Relat Res       Date:  2008-06-04       Impact factor: 4.176

6.  The Endo-Model(®) rotating hinge for rheumatoid knees : Functional results in primary and revision surgery.

Authors:  L Felli; M Coviello; M Alessio-Mazzola; M Cutolo
Journal:  Orthopade       Date:  2016-05       Impact factor: 1.087

7.  Revision knee arthroplasty with a rotating-hinge design in elderly patients with instability following total knee arthroplasty.

Authors:  E Carlos Rodríguez-Merchán; Primitivo Gómez-Cardero; Ángel Martínez-Lloreda
Journal:  J Clin Orthop Trauma       Date:  2014-11-20

8.  The impact of joint line restoration on functional results after hinged knee prosthesis.

Authors:  Serdar Yilmaz; Deniz Cankaya; Alper Deveci; Ahmet Firat; Bulent Ozkurt; Murat Bozkurt
Journal:  Indian J Orthop       Date:  2016 Mar-Apr       Impact factor: 1.251

9.  Posterior dislocation of the hinge-post extension in a rotating hinge total knee prosthesis.

Authors:  Givenchy Manzano; Ran Schwarzkopf
Journal:  Case Rep Orthop       Date:  2013-09-23

10.  Evaluation of stability of rotating hinge knee prostheses: a biomechanical study.

Authors:  Joerg Friesenbichler; Andreas Leithner; Mathias Glehr; Patrick Sadoghi; Werner Maurer-Ertl; Alexander Avian; Reinhard Windhager
Journal:  ISRN Orthop       Date:  2013-07-18