Literature DB >> 10751298

Can insertional torque predict screw loosening and related failures? An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion.

K Okuyama1, E Abe, T Suzuki, Y Tamura, M Chiba, K Sato.   

Abstract

STUDY
DESIGN: An investigation of the relation between intraoperative insertional torque of pedicle screws, bone mineral density of the vertebra, and development of screw loosening in vivo.
OBJECTIVES: To determine the usefulness of intraoperative measurement of the insertional torque of pedicle screws. SUMMARY OF BACKGROUND DATA: Some biomechanical studies have demonstrated that the insertional torque is highly correlated with bone mineral density and the stability of pedicle screws in vitro.
METHODS: Pedicle screw fixation was performed with posterior lumbar interbody fusion in 62 consecutive patients. The mean age of the patients at the time of surgery was 58 years. The insertional torque of pedicle screws was measured intraoperatively in all patients. The mean follow-up period was 2.7 years.
RESULTS: The mean insertional torque was 1.28 +/- 0.37 Nm in patients with screw loosening and 1.50 +/- 0. 40 Nm in patients without the problem. The mean insertional torque in patients with compression fractures in the upper vertebra adjacent to the fixed segment was 0.83 +/- 0.23 Nm. There was no significant difference between the mean insertional torque in patients with screw loosening and those without the condition. The mean insertional torque in patients without screw loosening was significantly greater than in patients with compression fractures (P < 0.01). A high correlation was found between insertional torque and bone mineral density (P < 0.01).
CONCLUSIONS: Although a high correlation was found between the insertional torque of pedicle screws and bone mineral density in vivo, the insertional torque could not objectively predict screw loosening.

Entities:  

Mesh:

Year:  2000        PMID: 10751298     DOI: 10.1097/00007632-200004010-00015

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  31 in total

1.  The use of computed tomography attenuation to evaluate osteoporosis following acute fractures of the thoracic and lumbar vertebra.

Authors:  Osa Emohare; Amanda Cagan; Robert Morgan; Rick Davis; Martin Asis; Julie Switzer; David W Polly
Journal:  Geriatr Orthop Surg Rehabil       Date:  2014-06

2.  Biomechanical evaluation of an expansive pedicle screw in calf vertebrae.

Authors:  Wei Lei; Zixiang Wu
Journal:  Eur Spine J       Date:  2005-04-30       Impact factor: 3.134

3.  A case of pedicle screw loosening treated by modified transpedicular screw augmentation with polymethylmethacrylate.

Authors:  Suk-Hyung Kang; Kyoung-Tae Kim; Seung Won Park; Young-Baeg Kim
Journal:  J Korean Neurosurg Soc       Date:  2011-01-31

4.  Development and validation of a quantitative method to assess pedicle screw loosening in posterior spine instrumentation on plain radiographs.

Authors:  Emin Aghayev; Nicolas Zullig; Peter Diel; Daniel Dietrich; Lorin M Benneker
Journal:  Eur Spine J       Date:  2013-10-31       Impact factor: 3.134

5.  Pullout strength after expandable polymethylmethacrylate transpedicular screw augmentation for pedicle screw loosening.

Authors:  Suk-Hyung Kang; Yong Jun Cho; Young-Baeg Kim; Seung Won Park
Journal:  J Korean Neurosurg Soc       Date:  2015-04-24

6.  The contribution of the cortical shell to pedicle screw fixation.

Authors:  Matthew Henry Pelletier; Nicky Bertollo; Darweesh Al-Khawaja; William Robert Walsh
Journal:  J Spine Surg       Date:  2017-06

7.  Insertion torque is not a good predictor of pedicle screw loosening after spinal instrumentation: a prospective study in 8 patients.

Authors:  Bengt Sandén; Claes Olerud; Sune Larsson; Yohan Robinson
Journal:  Patient Saf Surg       Date:  2010-09-03

Review 8.  Pedicle screw loosening: a clinically relevant complication?

Authors:  Fabio Galbusera; David Volkheimer; Sandra Reitmaier; Nikolaus Berger-Roscher; Annette Kienle; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2015-01-24       Impact factor: 3.134

9.  Experimental and clinical analysis of a posterolateral lumbar appendicular bone graft fusion.

Authors:  Jian-Wen Wang; Dong-Min Xiao; Hong Wu; Ming Ye; Xiong Li
Journal:  Int J Clin Exp Med       Date:  2015-12-15

10.  Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone.

Authors:  Wen-Chi Tsai; Po-Quang Chen; Tung-Wu Lu; Shing-Sheng Wu; Kao-Shang Shih; Shang-Chih Lin
Journal:  BMC Musculoskelet Disord       Date:  2009-04-30       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.