Comparison of impression materials for direct multi-implant impressions.

A G Wee1.   

Abstract

STATEMENT OF PROBLEM: Given that meticulous implant prosthodontic procedures are recommended to obtain the best possible intraoral fit, impression materials that are suitable for use with a direct impression technique warrant further investigation.
PURPOSE: This in vitro study compared the amount of torque required to rotate a square impression coping in an impression and evaluated the accuracy of solid implant casts fabricated from different impression materials.
MATERIAL AND METHODS: Two direct transfer implant impressions were made using 8 impression materials; the torque required to rotate an impression coping in the impressions was calculated. Ten direct transfer implant impressions were made from the master model and poured in a die stone (Resin Rock) for 3 of the 8 initial impression material groups. Linear distances between steel balls placed on each abutment replica were measured with a traveling microscope to determine distortion in the impression procedure for each group. Data were analyzed (P =.05) with ANOVA and Ryan-Einot-Gabriel-Welsch multiple range test for post hoc.
RESULTS: With a 1-way ANOVA, average torque values among the material groups differed significantly (P =.001). Polyether (medium consistency) was found to produce the highest overall torque values, followed by addition silicone (high consistency), and then polysulfide (medium consistency). Statistically significant difference was also found among the 3 material groups' mean absolute cast error using a 1-way ANOVA (P =.0086). Implant casts made from polyether (medium) or addition silicone (high) impressions were significantly more accurate than casts made from polysulfide medium impressions.
CONCLUSION: On the basis of the results of this study, the use of either polyether (medium) or addition silicone (high) impression is recommended for direct implant impressions.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10709042     DOI: 10.1016/s0022-3913(00)70136-3

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  18 in total

1.  Evaluation of accuracy of direct transfer snapon impression coping closed tray impression technique and direct transfer open tray impression technique: an in vitro study.

Authors:  T Balamurugan; P Manimaran
Journal:  J Indian Prosthodont Soc       Date:  2012-08-19

2.  Effect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study.

Authors:  Si-Hoon Jo; Kyoung-Il Kim; Jae-Min Seo; Kwang-Yeob Song; Ju-Mi Park; Seung-Geun Ahn
Journal:  J Adv Prosthodont       Date:  2010-12-31       Impact factor: 1.904

3.  In Vitro Comparative Evaluation of Different Types of Impression Trays and Impression Materials on the Accuracy of Open Tray Implant Impressions: A Pilot Study.

Authors:  Sonam Gupta; Aparna Ichalangod Narayan; Dhanasekar Balakrishnan
Journal:  Int J Dent       Date:  2017-02-27

4.  Accuracy of different impression materials in parallel and nonparallel implants.

Authors:  Mahroo Vojdani; Kianoosh Torabi; Elham Ansarifard
Journal:  Dent Res J (Isfahan)       Date:  2015 Jul-Aug

5.  A new portable vibrator for plaster pouring: effect on the marginal fit at cylinder-abutment.

Authors:  Pâmela Cândida Aires Ribas de Andrade; Leonardo Flores Luthi; Kyle Stanley; Antônio Carlos Cardoso
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

6.  Comparison of digital and conventional impression techniques: evaluation of patients' perception, treatment comfort, effectiveness and clinical outcomes.

Authors:  Emir Yuzbasioglu; Hanefi Kurt; Rana Turunc; Halenur Bilir
Journal:  BMC Oral Health       Date:  2014-01-30       Impact factor: 2.757

7.  Comparison of implant cast accuracy of multiple implant impression technique with different splinting materials: An in vitro study.

Authors:  Sunantha Selvaraj; Jayachandran Dorairaj; Jayashree Mohan; Paul Simon
Journal:  J Indian Prosthodont Soc       Date:  2016 Apr-Jun

8.  A comparative analysis of the accuracy of implant master casts fabricated from two different transfer impression techniques.

Authors:  Rupali Patil; Pankaj Kadam; Chetan Oswal; Seema Patil; Shweta Jajoo; Arati Gachake
Journal:  J Int Soc Prev Community Dent       Date:  2016 Mar-Apr

9.  Accuracy of Implant Position Transfer and Surface Detail Reproduction with Different Impression Materials and Techniques.

Authors:  Marzieh Alikhasi; Hakimeh Siadat; Elaheh Beyabanaki; Mohammad Javad Kharazifard
Journal:  J Dent (Tehran)       Date:  2015-10

10.  Changes in views on digital intraoral scanners among dental hygienists after training in digital impression taking.

Authors:  Hye-Ran Park; Ji-Man Park; Youn-Sic Chun; Kkot-Nim Lee; Minji Kim
Journal:  BMC Oral Health       Date:  2015-11-27       Impact factor: 2.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.