The effects of freeze drying and freeze drying additives on the prothrombin time and the international sensitivity index.

L Poller1, M Keown, S A Shepherd, C R Shiach, S Tabeart.   

Abstract

AIM: To determine whether freezing, freeze drying protective additives, or freeze drying of plasma samples from patients on coumarin treatment and from normal individuals affects prothrombin times or the international sensitivity index (ISI) calibration.
METHODS: The effect of the addition of the protective additives singly and combined on the prothrombin time of coumarin samples and normal samples before and after freeze drying was observed using high and low ISI reference thromboplastins. ISI values were also determined.
RESULTS: Freezing caused a prolongation of prothrombin time in the normal plasma samples with both reagents, which was significant with the low ISI human. Prolongation (non-significant) of the prothrombin time in coumarin plasma samples occurred with the human reagent only. Significant prolongation of normal prothrombin time by some of the protective additives before and after freeze drying was observed with both thromboplastins but to a greater extent with the human. Significant prolongation of prothrombin time in coumarin plasma samples was observed, but again was more marked with human thromboplastin. An approximate ISI was determined on the 20 coumarin samples. The only marked ISI change was with the WHO human thromboplastin after freeze drying of plasma, where a decrease from 0.95 to 0.90 was observed, corresponding to a marked prothrombin ratio increase.
CONCLUSIONS: Freeze drying additives and the freeze drying procedure prolong normal and coumarin prothrombin times, with low ISI thromboplastin. Less marked prolongations occurred with a high ISI rabbit reagent, coumarin samples showing more significant prolongations. Marked ISI change in freeze dried plasma was only recorded with the low ISI ECAA human reagent. Frozen normal plasma samples cannot be used with confidence for ISI calibrations.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10674031      PMCID: PMC501567          DOI: 10.1136/jcp.52.10.744

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  10 in total

1.  Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.

Authors:  K Clarke; D A Taberner; J M Thomson; J A Morris; L Poller
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

2.  The European Concerted Action on Anticoagulation (ECAA): field studies of coagulometer effects on the ISI of ECAA thromboplastins.

Authors:  L Poller; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Thromb Haemost       Date:  1998-10       Impact factor: 5.249

3.  Comparison of lyophilized plasmas with fresh plasmas for calibration of thromboplastin reagents in oral anticoagulant control.

Authors:  A M van den Besselaar
Journal:  Br J Haematol       Date:  1996-05       Impact factor: 6.998

4.  A comparison of artificially-depleted, lyophilized coumarin and fresh coumarin plasmas in thromboplastin calibration. European Concerted Action on Anticoagulation.

Authors:  L Poller; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Br J Haematol       Date:  1998-06       Impact factor: 6.998

5.  The European Concerted Action on Anticoagulation (ECAA) evaluation of a set of lyophilized normal plasmas to establish the normal prothrombin time for coagulometer systems.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Thromb Haemost       Date:  1998-01       Impact factor: 5.249

6.  Minimum lyophilized plasma requirement for ISI calibration. European Concerted Action on Anticoagulation.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Am J Clin Pathol       Date:  1998-02       Impact factor: 2.493

7.  European Concerted Action on Anticoagulation (ECAA)-the multicentre calibration of rabbit and human ECAA reference thromboplastins. Steering Committee.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Thromb Haemost       Date:  1996-12       Impact factor: 5.249

8.  The effect of sample size on fresh plasma thromboplastin ISI determination.

Authors:  L Poller; A M Van Den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Br J Haematol       Date:  1999-06       Impact factor: 6.998

9.  The reliability of the mean normal prothrombin time of fresh plasmas and of the normal value from a lyophilized 'normal' plasma in prothrombin ratio determination.

Authors:  L Poller; D Houghton; J Carroll
Journal:  Br J Haematol       Date:  1994-12       Impact factor: 6.998

10.  The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study.

Authors:  L Poller; D A Triplett; J Hirsh; J Carroll; K Clarke
Journal:  Am J Clin Pathol       Date:  1995-03       Impact factor: 2.493