| Literature DB >> 10655416 |
C Manegold1, C Krempe, H Jablonowski, L Kajala, M Dietrich, O Adams.
Abstract
We have comparatively evaluated Quantiplex version 3.0 and version 2. 0 on 133 plasma samples and a repetitive dilution series. Version 3. 0 yielded higher human immunodeficiency virus RNA values, and the ratio of version 3.0 results to version 2.0 results decreased from 3. 47 below 1,000 copies/ml to 1.97 above 50,000 copies/ml [linear regression, log (version 3.0) = 0.915 + 0.871 x log (version 2.0); r(2) = 0.952].Entities:
Mesh:
Substances:
Year: 2000 PMID: 10655416 PMCID: PMC86246
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948