Comparison of a human portable glucometer and an automated chemistry analyzer for measurement of blood glucose concentration in pet ferrets (Mustela putorius furo).

Noémie M Summa1, David Eshar1, Bridget Lee-Chow1, Sylvain Larrat1, Dorothy C Brown1.   

Abstract

This study compared blood glucose concentrations measured with a portable blood glucometer and a validated laboratory analyzer in venous blood samples of 20 pet ferrets (Mustela putorius furo). Correlation and agreement were evaluated with a Bland-Altman plot method and Lin's concordance correlation coefficient. Blood glucose concentrations measured with the laboratory analyzer and the glucometer ranged from 1.9 to 8.6 mmol/L and from 0.9 to 9.2 mmol/L, respectively. The glucometer had a poor agreement and correlation with the laboratory analyzer (bias, -0.13 mmol/L; level of agreement, -2.0 to 3.6 mmol/L, concordance correlation coefficient 0.665). The relative sensitivity and specificity of the portable blood glucometer for detection of hypoglycemia were 100% (95% CI: 66% to 100%) and 50% (95% CI: 20% to 80%), respectively. Positive and negative predictive values were 67% (95% CI: 39% to 87%) and 100% (95% CI: 46% to 100%), respectively. Based on these results, clinicians are advised to be cautious when considering the results from this handheld glucometer in pet ferrets, and blood glucose concentrations should be determined with a laboratory analyzer validated for this species.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25183894      PMCID: PMC4137928     

Source DB:  PubMed          Journal:  Can Vet J        ISSN: 0008-5286            Impact factor:   1.008


  27 in total

1.  Assessment of five portable blood glucose meters for use in cats.

Authors:  G Wess; C Reusch
Journal:  Am J Vet Res       Date:  2000-12       Impact factor: 1.156

2.  Measurement of glucose concentrations in rats: differences between glucose meter and plasma laboratory results.

Authors:  R Weitgasser; A M Davalli; G C Weir
Journal:  Diabetologia       Date:  1999-02       Impact factor: 10.122

3.  Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.

Authors:  Brad S Karon; Laurie Griesmann; Renee Scott; Sandra C Bryant; Jeffrey A Dubois; Terry L Shirey; Steven Presti; Paula J Santrach
Journal:  Diabetes Technol Ther       Date:  2008-04       Impact factor: 6.118

4.  Neoplastic diseases in ferrets: 574 cases (1968-1997).

Authors:  X Li; J G Fox; P A Padrid
Journal:  J Am Vet Med Assoc       Date:  1998-05-01       Impact factor: 1.936

5.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

6.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

7.  Consensus statement on self-monitoring of blood glucose.

Authors: 
Journal:  Diabetes Care       Date:  1987 Jan-Feb       Impact factor: 19.112

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Comparison of the Accu-Chek Aviva point-of-care glucometer with blood gas and laboratory methods of analysis of glucose measurement in equine emergency patients.

Authors:  A R Hollis; B L Dallap Schaer; R C Boston; P A Wilkins
Journal:  J Vet Intern Med       Date:  2008-07-11       Impact factor: 3.333

10.  Repeated blood glucose measures using a novel portable glucose meter.

Authors:  C Messier; P Kent
Journal:  Physiol Behav       Date:  1995-04
View more
  4 in total

1.  Accuracy of 5 Point-of-Care Glucometers in C57BL/6J Mice.

Authors:  Linnea A Morley; Thomas H Gomez; Julia L Goldman; Rene Flores; Mary A Robinson
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-01-01       Impact factor: 1.232

2.  Pneumatosis coli in a domestic ferret (Mustela putorius furo).

Authors:  Cédric B Larouche; Marion R Desmarchelier; Swan Specchi; Isabelle Langlois
Journal:  Can Vet J       Date:  2017-04       Impact factor: 1.008

3.  Obtaining accurate glucose measurements from wild animals under field conditions: comparing a hand held glucometer with a standard laboratory technique in grey seals.

Authors:  Kimberley A Bennett; Lucy M Turner; Sebastian Millward; Simon E W Moss; Ailsa J Hall
Journal:  Conserv Physiol       Date:  2017-02-27       Impact factor: 3.079

Review 4.  Medical and Surgical Emergencies in Ferrets.

Authors:  Nicola Di Girolamo; Paolo Selleri
Journal:  Vet Clin North Am Exot Anim Pract       Date:  2016-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.