Literature DB >> 10635193

Investigations into the failure of dental magnets.

M A Riley1, A J Williams, J D Speight, A D Walmsley, I R Harris.   

Abstract

PURPOSE: The aim of this study was to determine the mechanisms of failure of magnets used for denture retention.
MATERIALS AND METHODS: Magnets were retrieved from dentures that had failed after 4 to 18 months of clinical use. The magnets were sectioned and the failure modes were evaluated using optical or scanning electron microscopy.
RESULTS: The magnetic retention system investigated consists of a neodymium-iron-boron magnet sandwiched in a stainless steel keeper, capped with a titanium dome, and sealed with epoxy resin. The diffusion of moisture through the epoxy seal and the breakdown of the encapsulating material led to corrosion of the magnet. In both cases, a loss in retention occurred as the corrosion products weakened the overall magnetic attraction.
CONCLUSION: The failure of magnets may occur by either breakdown of the epoxy seal or breakdown of the encapsulating material. Both problems need to be addressed to prolong the life span of these magnets.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10635193

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  6 in total

1.  Natural frequency assessment of stability of root keeper magnetic devices.

Authors:  H M Huang; D Z Liu; Y Y Shiau; C Y Yeh; C T Lin; S Y Lee
Journal:  Med Biol Eng Comput       Date:  2004-05       Impact factor: 2.602

2.  A study on the changes in attractive force of magnetic attachments for overdenture.

Authors:  Han-Wool Leem; In-Ho Cho; Jong-Hyuk Lee; Yu-Sung Choi
Journal:  J Adv Prosthodont       Date:  2016-02-23       Impact factor: 1.904

3.  Attachment systems for mandibular implant overdentures: a systematic review.

Authors:  Ha-Young Kim; Jeong-Yol Lee; Sang-Wan Shin; S Ross Bryant
Journal:  J Adv Prosthodont       Date:  2012-11-29       Impact factor: 1.904

4.  Prosthetic rehabilitation of severe Siebert's Class III defect with modified Andrews bridge system.

Authors:  Manu Rathee; Neha Sikka; Sahil Jindal; Ashutosh Kaushik
Journal:  Contemp Clin Dent       Date:  2015-03

5.  Corrosion behaviours of the dental magnetic keeper complexes made by different alloys and methods.

Authors:  Min-Ke Wu; Ning Song; Fei Liu; Liang Kou; Xiao-Wen Lu; Min Wang; Hang Wang; Jie-Fei Shen
Journal:  Int J Oral Sci       Date:  2016-09-29       Impact factor: 6.344

6.  Cytotoxic effect of indigenously fabricated dental magnets for application in prosthodontics.

Authors:  Satyabodh Shesharaj Guttal; Ramesh K Nadiger; Pravinkumar Shetty
Journal:  J Indian Prosthodont Soc       Date:  2018 Jan-Mar
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.