Analysis of the detective quantum efficiency of a developmental detector for digital mammography.

M B Williams1, P U Simoni, L Smilowitz, M Stanton, W Phillips, A Stewart.   

Abstract

We are developing a modular detector for applications in full field digital mammography and for diagnostic breast imaging. The detector is based on a design that has been refined over the past decade for applications in x-ray crystallography [Kalata et al., Proc. SPIE 1345, 270-279 (1990); Phillips et al. ibid. 2009, 133-138 (1993), Phillips et al., Nucl. Instrum. Methods Phys. Rev. A 334, 621-630 (1993)]. The full field mammographic detector, currently undergoing clinical evaluation, is formed from a 19 cm x 28 cm phosphor screen, read out by a 2 x 3 array of butted charge-coupled device (CCD) modules. Each 2k x 2k CCD is optically coupled to the phosphor via a fiber optic taper with dimensions of 9.4 cm x 9.4cm at the phosphor. This paper describes the imaging performance of a two-module prototype, built using a similar design. In this paper we use cascaded linear systems analysis to develop a model for calculating the spatial frequency dependent noise power spectrum (NPS) and detective quantum efficiency (DQE) of the detector using the measured modulation transfer function (MTF). We compare results of the calculation with the measured NPS and DQE of the prototype. Calculated and measured DQEs are compared over a range of clinically relevant x-ray exposures and kVps. We find that for x-ray photon energies between 10 and 28 keV, the detector gain ranges between 2.5 and 3.7 CCD electrons per incident x-ray, or approximately 5-8 electrons per absorbed x ray. Using a Mo/Mo beam and acrylic phantom, over a detector entrance exposure range of approximately 10 to 80 mR, the volume under the measured 2-d NPS of the prototype detector is proportional to the x-ray exposure, indicating quantum limited performance. Substantial agreement between the calculated and measured values was obtained for the frequency and exposure dependent NPS and DQE over a range of tube voltage from 25 to 30 kVp.

Entities:  

Mesh:

Year:  1999        PMID: 10587208     DOI: 10.1118/1.598741

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Integrated model for estimating phosphor signal and noise transfer characteristics on medical images: application to CdPO3Cl:Mn phosphor screens.

Authors:  D Cavouras; I Kandarakis; G S Panayiotakis; C D Nomicos
Journal:  Med Biol Eng Comput       Date:  2002-05       Impact factor: 2.602

2.  Solid-state fluoroscopic imager for high-resolution angiography: parallel-cascaded linear systems analysis.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

3.  Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan; Steven K Onishi
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

4.  Dual-modality breast tomosynthesis.

Authors:  Mark B Williams; Patricia G Judy; Spencer Gunn; Stanislaw Majewski
Journal:  Radiology       Date:  2010-04       Impact factor: 11.105

5.  A method for the determination of the two-dimensional MTF of digital radiography systems using only the noise response.

Authors:  Andrew Kuhls-Gilcrist; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2010-03-23

6.  Comparison of two detector systems for cone beam CT small animal imaging - a preliminary study.

Authors:  Yang Meng; Chris C Shaw; Xinming Liu; Mustafa C Altunbas; Tianpeng Wang; Lingyun Chen; Shu-Ju Tu; S Cheenu Kappadath; Chao-Jen Lai
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-03-02

7.  New light-amplifier-based detector designs for high spatial resolution and high sensitivity CBCT mammography and fluoroscopy.

Authors:  Stephen Rudin; Andrew T Kuhls; Girijesh K Yadava; Gaurav C Josan; Ye Wu; Ravishankar N Chityala; Hussain S Rangwala; N Ciprian Ionita; Kenneth R Hoffmann; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-12-02

8.  Accurate MTF measurement in digital radiography using noise response.

Authors:  Andrew Kuhls-Gilcrist; Amit Jain; Daniel R Bednarek; Kenneth R Hoffmann; Stephen Rudin
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

9.  A theoretical and experimental evaluation of the microangiographic fluoroscope: A high-resolution region-of-interest x-ray imager.

Authors:  Amit Jain; D R Bednarek; Ciprian Ionita; S Rudin
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

10.  Progress in electron-multiplying CCD (EMCCD) based, high-resolution, high-sensitivity x-ray detector for fluoroscopy and radiography.

Authors:  Andrew T Kuhls; Girijesh Yadava; Vikas Patel; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2007
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.