Noninvasive glucose monitoring: comprehensive clinical results. Cygnus Research Team.

J A Tamada1, S Garg, L Jovanovic, K R Pitzer, S Fermi, R O Potts.   

Abstract

CONTEXT: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management.
OBJECTIVE: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements.
DESIGN: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study.
SETTING: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States. PARTICIPANTS: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin. MAIN OUTCOME MEASURES: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid.
RESULTS: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis.
CONCLUSION: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10573275     DOI: 10.1001/jama.282.19.1839

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  48 in total

Review 1.  Severe hypoglycemia in adults.

Authors:  Mary F Carroll; Mark R Burge; David S Schade
Journal:  Rev Endocr Metab Disord       Date:  2003-05       Impact factor: 6.514

Review 2.  Evolving approaches to intensive insulin therapy in type 1 diabetes: multiple daily injections, insulin pumps and new methods of monitoring.

Authors:  Elizabeth Stephens; Matthew Riddle
Journal:  Rev Endocr Metab Disord       Date:  2003-12       Impact factor: 6.514

3.  The Future of Glucose Monitoring.

Authors:  Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2016-02       Impact factor: 6.118

Review 4.  Role of self-monitoring of blood glucose in glycemic control.

Authors:  Andrew J Karter
Journal:  Endocr Pract       Date:  2006 Jan-Feb       Impact factor: 3.443

Review 5.  In vitro, in vivo and post explantation testing of glucose-detecting biosensors: current methods and recommendations.

Authors:  Heidi E Koschwanez; William M Reichert
Journal:  Biomaterials       Date:  2007-04-19       Impact factor: 12.479

6.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

Review 7.  Delays in minimally invasive continuous glucose monitoring devices: a review of current technology.

Authors:  D Barry Keenan; John J Mastrototaro; Gayane Voskanyan; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

Review 8.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

Review 9.  Transdermal drug delivery.

Authors:  Mark R Prausnitz; Robert Langer
Journal:  Nat Biotechnol       Date:  2008-11       Impact factor: 54.908

10.  Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.

Authors:  Tai-Ping Sun; Hsiu-Li Shieh; Congo Tak-Shing Ching; Yan-Dong Yao; Su-Hua Huang; Chia-Ming Liu; Wei-Hao Liu; Chung-Yuan Chen
Journal:  Int J Nanomedicine       Date:  2010-05-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.