Literature DB >> 10554342

Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: report of two cases.

S Aoyagi1, K Tayama, E Tayama, S Fukunaga, H Akashi, T Kawara, A Oryoji.   

Abstract

We report herein the cases of two patients who received replacement of aortic Bjork-Shiley Delrin (BSD) valves that had been implanted for over 20 years following the development of aortic regurgitation (AR) resulting from wear of a Delrin disc. Case 1 was a 61-year-old woman in whom echocardiography revealed marked left ventricular dilatation and moderate AR 23 years after an aortic valve replacement (AVR) with a 21-mm BSD valve. Case 2 was a 51-year-old woman in whom echocardiography revealed marked dilation of the right atrium and moderate AR 23 years and 8 months after an AVR with a 21-mm BSD valve, as well as a mitral valve replacement with a 3M Starr-Edwards (SE) ball valve and tricuspid annuloplasty. In both patients, the BSD valves were replaced with other mechanical valves at reoperation. Examination of the explanted BSD valves showed that the Delrin discs contained increases in the radial gaps and strut indentation grooves on the inflow and outflow surfaces. The type and magnitude of the wear on the Delrin discs in these valves were consistent with data reported in the literature for this valve design after similar implant duration.

Entities:  

Mesh:

Year:  1999        PMID: 10554342     DOI: 10.1007/s005950050656

Source DB:  PubMed          Journal:  Surg Today        ISSN: 0941-1291            Impact factor:   2.549


  15 in total

1.  Delrin as an occluder material.

Authors:  S H Teoh; R L Martin; S C Lim; K H Lee; C K Mok; W C Kwok
Journal:  ASAIO Trans       Date:  1990 Jul-Sep

2.  Transesophageal Doppler color flow imaging in the detection of native and Björk-Shiley mitral valve regurgitation.

Authors:  M A Taams; E J Gussenhoven; M K Cahalan; J R Roelandt; L A van Herwerden; H K The; N Bom; N de Jong
Journal:  J Am Coll Cardiol       Date:  1989-01       Impact factor: 24.094

3.  Variability of regurgitation in Björk-Shiley mitral valves and relationship to disc occluder design: an in vitro two-dimensional color-Doppler flow mapping study.

Authors:  P D Lindower; K C Dellsperger; B Johnson; K B Chandran; B F Vandenberg
Journal:  J Heart Valve Dis       Date:  1996-08

4.  Clinical experience with the Björk-Shiley Delrin tilting disc heart valve.

Authors:  K C Dellsperger; G L Grunkemeier; P D Stein
Journal:  J Heart Valve Dis       Date:  1996-08

5.  Surface and edge wear of Björk-Shiley Delrin heart valve discs.

Authors:  P A Engel; E S Clark; S H Park; H McKellop; K Thyagarajan; Z Zhao; P Campbell; D W Wieting
Journal:  J Heart Valve Dis       Date:  1996-08

6.  A comparison of wear patterns in vivo and in vitro for Björk-Shiley Delrin heart valve discs.

Authors:  K Thyagarajan; C Conlin; H L Milligan; D W Wieting
Journal:  J Heart Valve Dis       Date:  1996-08

7.  A new tilting disc valve prosthesis.

Authors:  V O Björk
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1969

8.  Delrin as implant material for valve occluders.

Authors:  V O Björk
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1972

9.  [Late malfunction of the Björk-Shiley valve prosthesis due to Delrin disc defacement].

Authors:  T Katsumata; T Shimakura; H Nakano; Y Shimamura; K Hoshino; T Harada; F Maejima; H Kohno; K Asakawa; A Yabuki
Journal:  Kyobu Geka       Date:  1993-11

10.  [Malfunction of the Björk-Shiley valve prosthesis due to disc defacement--report of two cases of successful reoperation].

Authors:  H Misumi; K Matsuo; K Ishihara; S Nakae; A Hashimoto; H Koyanagi
Journal:  Nihon Kyobu Geka Gakkai Zasshi       Date:  1989-03
View more
  1 in total

1.  Replacement of a Björk-Shiley Delrin aortic valve still functioning after 25 years.

Authors:  M Ismail Badak; Erdem Ali Ozkisacik; Mehmet Boga; Ugur Gurcun; Berent Discigil
Journal:  Tex Heart Inst J       Date:  2004
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.