Literature DB >> 2768926

[Malfunction of the Björk-Shiley valve prosthesis due to disc defacement--report of two cases of successful reoperation].

H Misumi, K Matsuo, K Ishihara, S Nakae, A Hashimoto, H Koyanagi.   

Abstract

Derlin-disc model Björk-Shiley valve prosthesis was reoperated due to the disc defacement. One had admitted cerebral embolism which was suggested due to the prosthetic valve malfunction of aortic position and the other had congestive heart failure due to the malfunction of mitral position. These patients had implanted Derlin-disc model Björk-Shiley valve prosthesis in the aortic and mitral position 10 and 11 years ago. One's cineangiocardiography showed mild aortic regurgitation and the other showed mitral regurgitation due to the malfunction of the prosthesis. The episode of cerebral embolism in one patient was suggested due to the malfunction of aortic prosthetic valve, and congestive heart failure in the other patient was due to the malfunction of mitral prosthetic valve. These prostheses which removed at the reoperation were observed with strat shape indentation and distinct of the disc which resulted the malfunction of the prostheses. In patient who replaced with Derlin-disc model Björk-Shiley valve prosthesis should be carefully followed up.

Entities:  

Mesh:

Year:  1989        PMID: 2768926

Source DB:  PubMed          Journal:  Nihon Kyobu Geka Gakkai Zasshi        ISSN: 0369-4739


  2 in total

1.  Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: report of two cases.

Authors:  S Aoyagi; K Tayama; E Tayama; S Fukunaga; H Akashi; T Kawara; A Oryoji
Journal:  Surg Today       Date:  1999       Impact factor: 2.549

2.  The 37-year durability of a Björk-Shiley Delrin-disc aortic valve prosthesis.

Authors:  Fabrizio Sansone; Edoardo Zingarelli; Guglielmo Mario Actis Dato; Giuseppe Punta; Roberto Flocco; Stefano del Ponte; Riccardo Casabona
Journal:  Tex Heart Inst J       Date:  2012
  2 in total