Literature DB >> 8230934

[Late malfunction of the Björk-Shiley valve prosthesis due to Delrin disc defacement].

T Katsumata1, T Shimakura, H Nakano, Y Shimamura, K Hoshino, T Harada, F Maejima, H Kohno, K Asakawa, A Yabuki.   

Abstract

A 55-year-old woman eighteen years after mitral valvular replacement with Delrin disc Björk-Shiley valve prosthesis underwent a reoperation of prosthetic valve replacement for the prosthetic malfunction due to disc defacement. The patient suffered from faintness and vertigo at rest. An echocardiographic examination showed a moderate mitral insufficiency with a normal disc movement. Precise examination on the removed prosthesis revealed accelerated defacement of disc margin which made the ring-disc clearance up to 0.35 mm and strut-shaped groove formation on the inlet surface of the disc occluder. These findings suggested a pronouncedly earlier disc wear than predicted by Björk and co-workers. We concluded, therefore, that a patient undergone a valve replacement with Delrin disc Björk-Shiley valve should be examined periodically by echocardiography even though being without any symptoms.

Entities:  

Mesh:

Year:  1993        PMID: 8230934

Source DB:  PubMed          Journal:  Kyobu Geka        ISSN: 0021-5252


  2 in total

1.  Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: report of two cases.

Authors:  S Aoyagi; K Tayama; E Tayama; S Fukunaga; H Akashi; T Kawara; A Oryoji
Journal:  Surg Today       Date:  1999       Impact factor: 2.549

2.  The 37-year durability of a Björk-Shiley Delrin-disc aortic valve prosthesis.

Authors:  Fabrizio Sansone; Edoardo Zingarelli; Guglielmo Mario Actis Dato; Giuseppe Punta; Roberto Flocco; Stefano del Ponte; Riccardo Casabona
Journal:  Tex Heart Inst J       Date:  2012