Detection of simulated interstitial lung disease and catheters with selenium, storage phosphor, and film-based radiography.

T M Bernhardt1, D Otto, G Reichel, K Ludwig, S Seifert, S Kropf, U Rapp-Bernhardt.   

Abstract

PURPOSE: To evaluate the diagnostic performance of storage phosphor and digital selenium radiography (DSR) with asymmetric and symmetric screen-film systems at different speeds in the detection of simulated interstitial lung disease and catheters.
MATERIALS AND METHODS: Patterns of simulated interstitial lung disease and catheters were superimposed over an anthropomorphic chest phantom. Hard-copy images were generated at DSR (200-, 400-, and 600-speed), storage phosphor radiography (200- and 400-speed), and asymmetric (400-speed) and symmetric (200- and 400-speed) screen-film imaging. Surface doses were measured, and receiver operating characteristic analyses were performed.
RESULTS: No statistically significant differences were found between the detector systems with the same speeds for each interstitial pattern. Significantly poorer results were found at 600-speed DSR than at 200-speed DSR. Detection of catheters and nodules over high-attenuation areas was significantly worse with the symmetric screen-film system than with the other detectors. The surface dose with the DSR system, without a grid, was about 50% less than that of the other detector systems, with grids, at the same speed.
CONCLUSION: No significant difference was found in the diagnostic performance at DSR, storage phosphor radiography, and film-based radiography for simulated interstitial lung disease at corresponding speeds; there was a reduction in the surface dose of about 50% with the 400-speed DSR system.

Entities:  

Mesh:

Year:  1999        PMID: 10551225     DOI: 10.1148/radiology.213.2.r99nv20445

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Low-dose CT screening using hybrid iterative reconstruction: confidence ratings of diagnoses of simulated lesions other than lung cancer.

Authors:  N Sakai; H Yabuuchi; M Kondo; Y Matsuo; T Kamitani; M Nagao; M Jinnouchi; M Yonezawa; T Kojima; Y Yano; H Honda
Journal:  Br J Radiol       Date:  2015-07-08       Impact factor: 3.039

2.  Detectability of simulated interstitial pneumonia on chest radiographs: comparison between irradiation side sampling indirect flat-panel detector and computed radiography.

Authors:  Y Yano; H Yabuuchi; A Kairada; N Tanaka; J Morishita; T Akasaka; Y Matsuo; T Kamitani; Y Yamasaki; M Nagao; M Sasaki
Journal:  Br J Radiol       Date:  2014-05-29       Impact factor: 3.039

3.  Modeling the performance characteristics of computed radiography (CR) systems.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  IEEE Trans Med Imaging       Date:  2010-03       Impact factor: 10.048

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.