Venous air in the bypass circuit: a source of arterial line emboli exacerbated by vacuum-assisted drainage.

T W Willcox1, S J Mitchell, D F Gorman.   

Abstract

BACKGROUND: Arterial emboli cause neurocognitive deficits in cardiac surgical patients. Carotid artery emboli, detected ultrasonically, have been observed after venous air entrainment into the cardiopulmonary bypass circuit. We investigated in vitro the extent to which venous air affected emboli detected in the arterial line downstream from a 40-microm filter.
METHODS: Using salvaged clinical cardiopulmonary bypass circuits, fixed volumes of air were introduced into the venous return line at unrestricted rates and at fixed rates using gravity venous drainage and vacuum-assisted venous drainage. Emboli counts were recorded distal to the arterial line filter using a 2-MHz pulsed-wave Doppler monitor. Emboli counts were similarly recorded after the introduction of carbon dioxide into the venous return line instead of air.
RESULTS: The number of emboli rose with increasing volumes of entrained venous air (p < 0.001), and there was an almost tenfold increase with vacuum-assisted venous drainage (p < 0.0001) compared with gravity venous drainage. Venous air was entrained at a significantly faster rate under vacuum-assisted venous drainage (p < 0.0001). When the entrainment rate of venous air was fixed, the difference in emboli numbers recorded for gravity and assisted venous drainage was not significant. There was a significant reduction in arterial line emboli when carbon dioxide rather than air was entrained under both vacuum-assisted and gravity drainage (p < 0.001).
CONCLUSIONS: Entrained venous air during cardiopulmonary bypass is a potential hazard, particularly during vacuum-assisted venous drainage. Every effort should be made to avoid venous air entrainment.

Entities:  

Mesh:

Year:  1999        PMID: 10543494     DOI: 10.1016/s0003-4975(99)00721-3

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  24 in total

1.  Microbubble generation and transmission of medtronic's affinity hardshell venous reservoir and collapsible venous reservoir bag: an in-vitro comparison.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2011-09

2.  Monitoring microemboli during cardiopulmonary bypass with the EDAC quantifier.

Authors:  John E Lynch; Christopher Wells; Tom Akers; Paul Frantz; Donna Garrett; M Lance Scott; Lisa Williamson; Barbara Agnew; John K Lynch
Journal:  J Extra Corpor Technol       Date:  2010-09

3.  Should Air Bubble Detectors Be Used to Quantify Microbubble Activity during Cardiopulmonary Bypass?

Authors:  Richard F Newland; Robert A Baker; Annette L Mazzone; Vijaykumar N Valiyapurayil
Journal:  J Extra Corpor Technol       Date:  2015-09

Review 4.  What blood pressure is appropriate for cardiopulmonary bypass and how to get it.

Authors:  Alan Merry
Journal:  J Extra Corpor Technol       Date:  2006-03

5.  Normalization of deviance: rocket science 101.

Authors:  Robert Groom
Journal:  J Extra Corpor Technol       Date:  2006-09

6.  Arterial bubbles from the venous line.

Authors:  Timothy W Willcox; Simon J Mitchell
Journal:  J Extra Corpor Technol       Date:  2006-09

Review 7.  Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass.

Authors:  Shigang Wang; Akif Undar
Journal:  J Extra Corpor Technol       Date:  2008-12

8.  Vacuum assist: angel or demon CON.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2013-06

Review 9.  Clinical microsystems: a critical framework for crossing the quality chasm.

Authors:  Donald S Likosky
Journal:  J Extra Corpor Technol       Date:  2014-03

10.  Arterial line filters ranked for gaseous micro-emboli separation performance: an in vitro study.

Authors:  Jeffrey B Riley
Journal:  J Extra Corpor Technol       Date:  2008-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.