UNLABELLED: This study compared the multiring detector (Ring-PET) and the dual-head coincidence imaging system (DH-PET) for staging/ restaging neoplastic patients before or after surgery or radiochemotherapy.
METHODS: Seventy patients with suspected tumor recurrence or metastatic dissemination received an intravenous dose of 18F-fluorodeoxyglucose (FDG) under overnight fasting and were studied in sequence with a dedicated positron emission tomograph with Ring-PET and a DH-PET. Ring-PET studies were performed 45-75 min postinjection and were followed by a DH-PET scan approximately 3 h postinjection. Number and location of the hypermetabolic lesions detected on DH-PET and Ring-PET reconstructed images were blindly assessed by three independent observers.
RESULTS: DH-PET identified all 14 head lesions detected by Ring-PET, 53 of 63 thoracic lesions and 36 of 45 abdominal lesions. Of the 19 lesions not identified by DH-PET, 6 were smaller than 10 mm, 8 were between 10 and 15 mm and 1 was 18 mm; dimensions of 4 bone lesions were not available. A concordant restaging, based on location and number of lesions detected, was found in all 14 patients with head tumors, in 28 of 30 patients with thoracic tumors and in 24 of 26 patients with abdominal tumors.
CONCLUSION: We found a good agreement between Ring-PET and DH-PET assessment of oncologic patients in detecting hypermetabolic lesions > or = 10-15 mm.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10520700

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  2 in total

Review 1.  Impact of technology on the utilisation of positron emission tomography in lymphoma: current and future perspectives.

Authors:  D Visvikis; P J Ell
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-05-13       Impact factor: 9.236

2.  [18F]FDG in recurrent breast cancer: diagnostic performances, clinical impact and relevance of induced changes in management.

Authors:  Dany Grahek; Françoise Montravers; Khaldoun Kerrou; Nicolas Aide; Jean-Pierre Lotz; Jean-Noël Talbot
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-02       Impact factor: 9.236