Evaluation of a rapid whole blood ELISA for quantification of troponin I in patients with acute chest pain.

C Heeschen1, B U Goldmann, L Langenbrink, G Matschuck, C W Hamm.   

Abstract

BACKGROUND: Troponin I (cTnI) provides important prognostic information in patients with chest pain. We wished to evaluate a rapid, whole-blood analyzer for quantitative point-of-care testing.
METHODS: A quantitative point-of-care test system (Stratus CS((R)); Dade-Behring) for cTnI with an incorporated centrifuge was evaluated in 412 patients with chest pain less than 12 h.
RESULTS: Results were available within 15 min. CVs were 4.5% at 0.1 microgram/L, 4.2% at 0.25 microgram/L, and 6.5% at 0.82 microgram/L. The detection limit was 0. 01 microgram/L. The 97.5% percentile in a healthy population was 0.08 microgram/L. Based on ROC curve analysis, a threshold of 0.15 microgram/L was calculated for the detection of acute myocardial infarction (AMI). With it, sensitivity for the detection of patients with AMI (n = 62) was 63% at arrival and 98% after 4 h (Stratus II((R)), 48% and 85%, respectively; P <0.01). In 42% of patients with unstable angina (n = 121), cTnI was >/=0.08 microgram/L (Stratus II, 28%; P <0. 01). During 30 days, death or AMI occurred in 25.5% of these cTnI-positive vs 2.9% of cTnI-negative patients (Stratus II, 29.4% vs 5.8%).
CONCLUSION: The Stratus CS provided better analytical performance and comparable or better prognostic information than the Stratus II.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10508126

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

1.  The reporting of cardiac troponin.

Authors:  Peter E Hickman; Gus Koerbin; Jill Tate
Journal:  Clin Biochem Rev       Date:  2003-11

Review 2.  Effectiveness of practices for improving the diagnostic accuracy of Non ST Elevation Myocardial Infarction in the Emergency Department: A Laboratory Medicine Best Practices™ systematic review.

Authors:  Christopher Layfield; John Rose; Aaron Alford; Susan R Snyder; Fred S Apple; Farah M Chowdhury; Michael C Kontos; L Kristin Newby; Alan B Storrow; Milenko Tanasijevic; Elizabeth Leibach; Edward B Liebow; Robert H Christenson
Journal:  Clin Biochem       Date:  2015-02-07       Impact factor: 3.281

3.  Discriminative potential of ion mobility spectrometry for the detection of fentanyl and fentanyl analogues relative to confounding environmental interferents.

Authors:  Thomas P Forbes; Jeffrey Lawrence; Jennifer R Verkouteren; R Michael Verkouteren
Journal:  Analyst       Date:  2019-10-03       Impact factor: 4.616

4.  Evaluation of Stratus CS stat fluorimetric analyser for measurement of cardiac markers Troponin I (cTnI), creatine kinase MB (CK-MB), and myoglobin.

Authors:  B Bénéteau-Burnat; B Baudin; M Vaubourdolle
Journal:  J Clin Lab Anal       Date:  2001       Impact factor: 2.352

5.  Non-linear transformation of enzyme-linked immunosorbent assay (ELISA) measurements allows usage of linear models for data analysis.

Authors:  Thomas M Lange; Maria Rotärmel; Dominik Müller; Gregory S Mahone; Friedrich Kopisch-Obuch; Harald Keunecke; Armin O Schmitt
Journal:  Virol J       Date:  2022-05-18       Impact factor: 5.913

6.  Ion mobility spectrometry nuisance alarm threshold analysis for illicit narcotics based on environmental background and a ROC-curve approach.

Authors:  Thomas P Forbes; Marcela Najarro
Journal:  Analyst       Date:  2016-05-20       Impact factor: 4.616

7.  Magnetized carbon nanotubes for visual detection of proteins directly in whole blood.

Authors:  Yan Huang; Yongqiang Wen; Kwaku Baryeh; Sunitha Takalkar; Michelle Lund; Xueji Zhang; Guodong Liu
Journal:  Anal Chim Acta       Date:  2017-09-27       Impact factor: 6.558

8.  Growth-differentiation factor-15 for early risk stratification in patients with acute chest pain.

Authors:  Kai M Eggers; Tibor Kempf; Tim Allhoff; Bertil Lindahl; Lars Wallentin; Kai C Wollert
Journal:  Eur Heart J       Date:  2008-07-29       Impact factor: 29.983

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.