In vitro flow properties of glaucoma implant devices.

D L Eisenberg1, E Y Koo, G Hafner, J S Schuman.   

Abstract

PURPOSE: To develop a protocol for testing glaucoma implant devices and to use this protocol to characterize devices currently available.
METHODS: The following devices were obtained: Ahmed Glaucoma Valve Implant, Baerveldt Implant, Krupin Eye Valve, Joseph Valve, and OptiMed Glaucoma Pressure Regulator. Pressure per unit time was measured in real-time during ramped pressure perfusion of the implant devices with an open-manometer system. Each device was measured during a 10 minute interval while a constant flow of fluid was pumped into the system. The open manometer allowed the resistance of the devices to generate a pressure head which was monitored by computer. Four independent runs were averaged for each device.
RESULTS: The Ahmed devices demonstrated reliable valve performance with a mean opening pressure of 13.65 mm Hg, a facility flow of 1.2 microL/min/mm Hg and a closing pressure of 6.1 mm Hg. The Baerveldt devices had a mean facility flow of 7.56 microL/min/mm Hg. The Joseph devices had the most variable performance, opening from 2.05 to 6.21 mm Hg. Two of 4 Joseph devices had high facility of flow (5 microL/min/mm Hg). The Krupin devices had a mean opening pressure of 6.25 mm Hg in 3 of 4 devices. The remaining Krupin device did not exhibit valve behavior. The OptiMed device did not reveal an opening pressure, and had a mean facility of flow of 7.08 microL/min/mm Hg for high flow and 6.20 microL/min/mm Hg for low flow.
CONCLUSION: An objective test protocol for glaucoma implant devices was demonstrated. All implant devices had high facility of flow compared to the normal eye and can be grouped into very high facility (Baerveldt and OptiMed), and high facility (Ahmed, Joseph and Krupin). Only the Ahmed device had consistent valve behavior. None of the devices created enough resistance to explain long-term clinical failure of glaucoma drainage device surgery.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10507569

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers        ISSN: 1082-3069


  9 in total

1.  [Ahmed glaucoma valve via pars plana access. Long-term results of implantation for therapy refractive glaucoma].

Authors:  B Mazinani; H Schwarzer; A Willkomm; A Weinberger; N Plange; P Walter; G Rössler
Journal:  Ophthalmologe       Date:  2013-06       Impact factor: 1.059

2.  Control and optimisation of fluid flow in glaucoma drainage device surgery.

Authors:  Kin Sheng Lim
Journal:  Eye (Lond)       Date:  2018-01-19       Impact factor: 3.775

3.  Evaluation of Changes in Refractive Errors Before and After Trabeculectomy and Shunt Surgery.

Authors:  Farzaneh Dehghanian Nasrabadi; Hassan Hashemi; Mohammad Aghazadeh Amiri; Ghasem Fakhraie; Alireza Akbarzadeh Baghban; Ebrahim Jafarzadehpour; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  Clin Optom (Auckl)       Date:  2020-07-29

4.  Risk factors for flat anterior chamber after glaucoma filtration surgery.

Authors:  Xiang-Ji Li; Richard Filek; Xiang-Ge He; Wei Wang; Hong Liu; Lian He; Yu-Rong Tang; Lin Xie
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

5.  [The Ahmed glaucoma valve. Medium term results].

Authors:  K Hille; M Rossi; A Hille; K W Ruprecht
Journal:  Ophthalmologe       Date:  2005-02       Impact factor: 1.059

6.  The Inflammatory and Foreign Body Reaction of Polymethyl Methacrylate Glaucoma Drainage Device in the Rabbit Eye.

Authors:  Virna D Oktariana Asrory; Ratna Sitompul; Widya Artini; Sri Estuningsih; Deni Noviana; William H Morgan
Journal:  Transl Vis Sci Technol       Date:  2020-02-18       Impact factor: 3.283

7.  Is Ahmed Glaucoma Valve Consistent in Performance?

Authors:  Nikhil S Choudhari; Swathi Vallabh Badakere; Ashutosh Richhariya; Sai Naga Sri Harsha Chittajallu; Sirisha Senthil; Chandra Sekhar Garudadri
Journal:  Transl Vis Sci Technol       Date:  2018-06-22       Impact factor: 3.283

8.  A Multi-Lumen Glaucoma Stent With Titratable Pressure-Lowering Effect: A Proof of Concept Study.

Authors:  Alexander Sverstad; Eirik Albert Torheim; Øystein Kalsnes Jørstad
Journal:  Transl Vis Sci Technol       Date:  2022-09-01       Impact factor: 3.048

9.  A One-Year Follow-Up of Two Ahmed Glaucoma Valve Models (S2 and FP7) for Refractory Glaucoma: A Prospective Randomized Trial.

Authors:  Khaled Hamdi Elbaklish; Wael Adel Gomaa
Journal:  Clin Ophthalmol       Date:  2020-03-04
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.