Literature DB >> 10461300

Five-year follow-up of a bipolar steroid-eluting ventricular pacing lead.

B Schwaab1, G Fröhlig, M Berg, H Schwerdt, H Schieffer.   

Abstract

Steroid-eluting pacing leads are known to attenuate the threshold peaking early after implantation. Long-term performance, however, is not yet settled. The lead design tested in this prospective study combines a 5.8-mm2 tip of microporous platinum-iridium with elution of 1.0 mg of dexamethasone sodium phosphate and tines for passive fixation (model 5024, Medtronic Inc.). In 50 patients (mean age 69 +/- 10 years), the electrode was implanted in the right ventricular apex. Follow-up was performed on days 0, 2, 5, 10, 28, 90, 180 and every 6 months thereafter for 5-years postimplant. At each visit, pacing thresholds were determined as pulse duration (ms) at 1.0 V and as the minimum charge (microC) delivered for capture. Lead impedance (omega) was telemetered at 2.5 V-0.50 ms, and sensing thresholds (mV) were measured in triplicate using the automatic sensing threshold algorithm of the pacemaker implanted (model 294-03, Intermedics Inc.). On the day of implantation, mean values were 0.10 +/- 0.03 ms, 0.12 +/- 0.03 microC, 758 +/- 131 omega, and 13.1 +/- 1.8 mV, respectively. Beyond 1-year postimplant, pacing thresholds did not vary significantly. Sensing thresholds and lead impedance values were stable during long-term follow-up. Five years after implantation, mean values were 0.23 +/- 0.11 ms, 0.24 +/- 0.07 microC, 670 +/- 139 omega, and 11.6 +/- 3.1 mV for pulse width and charge threshold, lead impedance, and sensing threshold, respectively, and all leads captured at 1.0 V with the longest pulse duration available (1.50 ms). It is concluded that the bipolar steroid-eluting tined ventricular lead showed stable stimulation thresholds, lead impedance values, and sensing thresholds for 5 years after implantation.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10461300     DOI: 10.1111/j.1540-8159.1999.tb00604.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Effect of left ventricular function on long-term left ventricular pacing and sensing threshold.

Authors:  Hung-Fat Tse; Cannas Yu; Vince E Paul; Giuseppe Boriani; Andreas Schuchert; Juan Leal del Ojo; Klaus Malinowski; Jean-Jacques Blanc; Chu-Pak Lau
Journal:  J Interv Card Electrophysiol       Date:  2003-08       Impact factor: 1.900

2.  Vascular endothelial growth factor and dexamethasone release from nonfouling sensor coatings affect the foreign body response.

Authors:  L W Norton; H E Koschwanez; N A Wisniewski; B Klitzman; W M Reichert
Journal:  J Biomed Mater Res A       Date:  2007-06-15       Impact factor: 4.396

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.