| Literature DB >> 10385287 |
Abstract
We examine the spatial resolution and variance properties of PET images reconstructed using maximum a posteriori (MAP) or penalized-likelihood methods. Resolution is characterized by the contrast recovery coefficient (CRC) of the local impulse response. Simplified approximate expressions are derived for the local impulse response CRC's and variances for each voxel. Using these results we propose a practical scheme for selecting spatially variant smoothing parameters to optimize lesion detectability through maximization of the local CRC-to-noise ratio in the reconstructed image.Mesh:
Year: 1999 PMID: 10385287 DOI: 10.1109/42.768839
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048