Metal debris from titanium spinal implants.

J C Wang1, W D Yu, H S Sandhu, F Betts, S Bhuta, R B Delamarter.   

Abstract

STUDY
DESIGN: A prospective study of tissue surrounding spinal instrumentation was performed using histologic and chemical analysis.
OBJECTIVES: To identify and quantify the amount of metal debris generated by titanium pedicle screw instrumentation and to evaluate the histologic response in the spinal tissues. SUMMARY OF BACKGROUND DATA: Microscopic metal particles from the soft tissue surrounding joint arthroplasties have been shown to activate a macrophage response that leads to bone resorption and increased inflammation. The use of titanium spinal implants for spine surgery projects the possibility of generating wear debris in the spine.
METHODS: Nine patients with titanium instrumentation from a prior lumbar decompression and fusion procedure who were undergoing reoperation were entered into this study. Tissue samples were collected from areas near the pedicle screw-rod junction, the scar tissue overlying the dura, and the pedicle screw holes. Metal levels for titanium were determined by electrothermal atomic absorption spectroscopy, and histologic analysis was performed by light and electron microscopy.
RESULTS: Tissue concentrations of titanium were highest in patients with a pseudarthrosis (30.36 micrograms/g of dry tissue). Patients with a solid fusion had low concentrations of titanium (0.586 microgram/g of dry tissue). Standard light microscopy identified metal particles in the soft tissues. Transmission electron microscopy demonstrated macrophages with numerous secondary lysosomes containing electron-dense bodies and collagenous stroma with electron-dense rod-like profiles consistent with metal debris.
CONCLUSIONS: Wear debris is generated by the use of titanium spinal instrumentation in patients with a pseudarthrosis. These particles activate a macrophage cellular response in the spinal tissues similar to that seen in surrounding joint prostheses. Patients with a solid spinal fusion have negligible levels of particulate matter.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10327512     DOI: 10.1097/00007632-199905010-00011

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  15 in total

1.  Quantifying subtle but persistent peri-spine inflammation in vivo to submicron cobalt-chromium alloy particles.

Authors:  Nadim James Hallab; Frank W Chan; Megan L Harper
Journal:  Eur Spine J       Date:  2012-03-10       Impact factor: 3.134

2.  Serum titanium, niobium and aluminium levels two years following instrumented spinal fusion in children: does implant surface area predict serum metal ion levels?

Authors:  Thomas P Cundy; William J Cundy; Georgia Antoniou; Leanne M Sutherland; Brian J C Freeman; Peter J Cundy
Journal:  Eur Spine J       Date:  2014-11       Impact factor: 3.134

3.  Comparative analysis between shape memory alloy-based correction and traditional correction technique in pedicle screws constructs for treating severe scoliosis.

Authors:  Yan Wang; Guoquan Zheng; Xuesong Zhang; Yonggang Zhang; Songhua Xiao; Zheng Wang
Journal:  Eur Spine J       Date:  2009-11-10       Impact factor: 3.134

4.  The spinal cord dura mater reaction to nitinol and titanium alloy particles: a 1-year study in rabbits.

Authors:  Souad Rhalmi; Sylvie Charette; Michel Assad; Christine Coillard; Charles H Rivard
Journal:  Eur Spine J       Date:  2007-03-03       Impact factor: 3.134

5.  Electron Microprobe Analysis and Tissue Reaction around Titanium Alloy Spinal Implants.

Authors:  Hee-Dong Kim; Ki-Soo Kim; Sung-Chan Ki; Yong-Soo Choi
Journal:  Asian Spine J       Date:  2007-06-30

6.  PEEK versus titanium-coated PEEK cervical cages: fusion rate.

Authors:  Bartosz Godlewski; Adam Bebenek; Maciej Dominiak; Grzegorz Karpinski; Piotr Cieslik; Tomasz Pawelczyk
Journal:  Acta Neurochir (Wien)       Date:  2022-04-26       Impact factor: 2.216

7.  Are Serum Ion Levels Elevated in Pediatric Patients With Metal Implants?

Authors:  Smitha E Mathew; Yong Xie; Leila Bagheri; Liam E Claton; Lin Chu; Amr Badreldin; Matthew P Abdel; Andre J van Wijnen; Geoffrey F Haft; Todd A Milbrandt; A Noelle Larson
Journal:  J Pediatr Orthop       Date:  2022-03-01       Impact factor: 2.324

Review 8.  Metallic Implants Used in Lumbar Interbody Fusion.

Authors:  Jakub Litak; Michał Szymoniuk; Wojciech Czyżewski; Zofia Hoffman; Joanna Litak; Leon Sakwa; Piotr Kamieniak
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

9.  Temporary use of shape memory spinal rod in the treatment of scoliosis.

Authors:  Yan Wang; Guoquan Zheng; Xuesong Zhang; Yonggang Zhang; Songhua Xiao; Zheng Wang
Journal:  Eur Spine J       Date:  2010-07-14       Impact factor: 3.134

10.  The development of whole blood titanium levels after instrumented spinal fusion - is there a correlation between the number of fused segments and titanium levels?

Authors:  Ingmar Ipach; Ralf Schäfer; Falk Mittag; Carmen Leichtle; Petra Wolf; Torsten Kluba
Journal:  BMC Musculoskelet Disord       Date:  2012-08-27       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.