Literature DB >> 10220194

Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty.

C H Huang1, C Y Yang, C K Cheng.   

Abstract

Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty (TKA) has not been well reported before. A 63-year-old man with osteoarthritis of the right knee underwent TKA with a New Jersey LCS Knee, with cementing on the tibia and patella but not on the femoral component. After 42 months, in addition to wearing of polyethylene of the tibia and patella, severe osteonecrosis of the medial femoral condyle was noted. Osteonecrosis caused loss of osseous support of the medial flange of the femoral component, and the bone ingrowth of the central and lateral flange to the distal femur was so good that it overcame the yield stress of the metal of the femoral component and caused fracture of the femoral component. The osteolytic area was filled with autogenous iliac bone, and a new femoral component was inserted and cemented. The patient's condition became satisfactory with relief of pain. Although uncommon, fracture of the femoral component does occur associated with polyethylene wear and osteolysis.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10220194     DOI: 10.1016/s0883-5403(99)90066-9

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  9 in total

1.  A rare cause for knee pain: fracture of the femoral component after TKR. A case report.

Authors:  C Luring; L Perlick; T Schubert; M Tingart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-10-06       Impact factor: 4.342

2.  Revision total knee arthroplasty for major osteolysis.

Authors:  R Stephen J Burnett; James A Keeney; William J Maloney; John C Clohisy
Journal:  Iowa Orthop J       Date:  2009

3.  Fracture at the stem-condylar junction of a modular femoral prosthesis in a varus-valgus constrained total knee arthroplasty.

Authors:  D D Nikolopoulos; I G Polyzois; E A Magnissalis; P F Bernard; I V Michos
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-08-25       Impact factor: 4.342

4.  Fatigue fracture of the hinge pin in a semi-constrained total knee arthroplasty: a case report.

Authors:  H R Springorum; C Luring; J Beckmann; J Schaumburger; J Grifka; M Tingart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-06-19       Impact factor: 4.342

5.  Tibial baseplate fracture associated with polyethylene wear and osteolysis after total knee arthroplasty.

Authors:  Deniz Cankaya; Serdar Yilmaz; Alper Deveci; Bulent Ozkurt
Journal:  Int J Surg Case Rep       Date:  2013-12-21

6.  Computational study on the effect of malalignment of the tibial component on the biomechanics of total knee arthroplasty: A Finite Element Analysis.

Authors:  D-S Suh; K-T Kang; J Son; O-R Kwon; C Baek; Y-G Koh
Journal:  Bone Joint Res       Date:  2017-11       Impact factor: 5.853

7.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

8.  FRACTURE OF THE TIBIAL COMPONENT IN TOTAL KNEE ARTHROPLASTY: REPORT ON TWO CASES.

Authors:  Idemar Monteiro da Palma; Rodrigo Pires E Albuquerque; João Maurício Barretto
Journal:  Rev Bras Ortop       Date:  2015-12-08

9.  The mechanical response of a polyetheretherketone femoral knee implant under a deep squatting loading condition.

Authors:  Lennert de Ruiter; Dennis Janssen; Adam Briscoe; Nico Verdonschot
Journal:  Proc Inst Mech Eng H       Date:  2017-11-04       Impact factor: 1.617

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.