Significant safety advantages gained with an improved pressure-regulated blood pump.

J P Montoya1, S I Merz, R H Bartlett.   

Abstract

A prototype of a non-occlusive pressure-regulated blood pump (M-pump) was evaluated in-vitro for safety in a comparative study with the roller and centrifugal pumps. The M-pump consists of collapsible tubing of unique design wrapped under tension around a rotor without a stator. The prototype M-pumps were tested in vitro to evaluate performance with respect to flow/ pressure characteristics, hemolysis, bubble generation (cavitation) and durability. The M-pump and centrifugal pump did not overpressurize at any RPM when the pump outlet was occluded, but the roller pump reached pressures in excess of 1000 mmHg. The M-pump did not generate negative pressures upon occlusion of the inlet, whereas the roller and centrifugal pumps produced near-vacuum pressures. Furthermore, the M-pump was unable to empty a blood reservoir when the height of the pump inlet was placed slightly above the reservoir outlet. The levels of microbubbles in the M-pump were significantly lower than the roller and centrifugal pumps upon sudden restriction of the pump inlet as determined with an ultrasonic bubble detector. The results of our in-vitro evaluation of the M-pump have shown it to have lower hemolysis than the centrifugal pump and lower or comparable hemolysis to roller pumps at flowrates of 0.1, 0.5, 4.0 and 6 L/min. We determined that the M-pump design possesses important safety advantages over roller and centrifugal pumps for cardiopulmonary bypass applications.

Entities:  

Mesh:

Year:  1996        PMID: 10160447

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  7 in total

1.  Successful Porcine Renal Transplantation After 60 Minutes of Donor Warm Ischemia: Extracorporeal Perfusion and Thrombolytics.

Authors:  David S Demos; Amit Iyengar; Benjamin S Bryner; Brian W Gray; Hayley R Hoffman; Marie S Cornell; John E Wilkinson; Daniel E Mazur; Robert H Bartlett; Jeffrey D Punch; Alvaro Rojas-Peña
Journal:  ASAIO J       Date:  2015 Jul-Aug       Impact factor: 2.872

2.  Preliminary in vivo testing of a novel pump for short-term extracorporeal life support.

Authors:  David J Spurlock; Kelly Koch; Daniel E Mazur; Emilia M Fracz; Robert H Bartlett; Jonathan W Haft
Journal:  Ann Thorac Surg       Date:  2011-11-23       Impact factor: 4.330

3.  In vitro testing of a novel blood pump designed for temporary extracorporeal support.

Authors:  David J Spurlock; David N Ranney; Emilia M Fracz; Daniel E Mazur; R H Bartlet; Jonathan W Haft
Journal:  ASAIO J       Date:  2012 Mar-Apr       Impact factor: 2.872

4.  A novel, low cost, disposable, pediatric pulsatile rotary ventricular pump for cardiac surgery that provides a physiological flow pattern.

Authors:  Daniel E Mazur; Kathryn R Osterholzer; John M Toomasian; Scott I Merz
Journal:  ASAIO J       Date:  2008 Sep-Oct       Impact factor: 2.872

5.  Post-arterial filter gaseous microemboli activity of five integral cardiotomy reservoirs during venting: an in vitro study.

Authors:  Gerard J Myers; Cheri Voorhees; Rob Haynes; Bob Eke
Journal:  J Extra Corpor Technol       Date:  2009-03

6.  A novel rotary pulsatile flow pump for cardiopulmonary bypass.

Authors:  Nicholas R Teman; Daniel E Mazur; John Toomasian; Emilia Jahangir; Fares Alghanem; Marcus Goudie; Alvaro Rojas-Peña; Jonathan W Haft
Journal:  ASAIO J       Date:  2014 May-Jun       Impact factor: 2.872

7.  Tidal Flow Perfusion for the Artificial Placenta: A Paradigm Shift.

Authors:  Jacqueline C Kading; Mark W Langley; Gergely Lautner; Mark M P Jeakle; John M Toomasian; Taylor L Fegan; Rachel A Pfannes; Samantha C Toor; Mary A Reiber; Peter R Kordell; Marie S Cornell; Robert H Bartlett; Alvaro Rojas-Pena; George B Mychaliska
Journal:  ASAIO J       Date:  2020-07       Impact factor: 3.826

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.