Literature DB >> 9930449

In vitro testing of bioprostheses: influence of mechanical stresses and lipids on calcification.

M Deiwick1, B Glasmacher, H A Baba, N Roeder, H Reul, G von Bally, H H Scheld.   

Abstract

BACKGROUND: Structural valve deterioration of bioprostheses is mainly caused by the progressive development of calcification. Mechanical stresses or lipid deposits in porcine aortic leaflets have been proposed as major factors contributing to the calcification process.
METHODS: A new test protocol consisting of nondestructive holographic interferometry, which allows a quantitative deformation analysis of heart valves, and accelerated dynamic in vitro calcification was used. The rapid calcification fluid contained a final combined calcium and phosphorus concentration of 130 (mg/dL)2 in barbital buffer solution. The calcification of 32 bioprostheses donated by different manufacturers (SJM Bioimplant, Biocor standard, Biocor No-React, Carpentier-Edwards SAV, Bravo, pericardial prototype) was assessed after up to 25 x 10(6) cycles by microradiography and the areas of calcification were compared with the holographic interferograms. The distribution of lipid droplets of four porcine prostheses were visualized by Sudan III stain before the calcification process.
RESULTS: Most of the tested bioprostheses had areas presenting with stress concentrations, and the dynamic in vitro testing resulted in leaflet calcification corresponding to the holographic irregularities. A strong correlation between calcification and stress distribution or lipid accumulation was found (r = 0.72; r = 0.81, respectively). After 19 x 10(6) cycles, the Carpentier-Edwards SAV and the pericardial valves had significantly less calcification than other prostheses tested (p = 0.003), but the variation among individual prostheses from the same manufacturer was even more pronounced.
CONCLUSIONS: Mechanical stresses or lipid accumulation seems to play an important role in the calcification process of bioprostheses. Quality control of bioprosthetic valves using holographic interferometry has the potential to predict calcification before implantation.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9930449     DOI: 10.1016/s0003-4975(98)01125-4

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  8 in total

1.  Effects of stent sizing on endothelial and vessel wall stress: potential mechanisms for in-stent restenosis.

Authors:  Henry Y Chen; James Hermiller; Anjan K Sinha; Michael Sturek; Luoding Zhu; Ghassan S Kassab
Journal:  J Appl Physiol (1985)       Date:  2009-03-19

2.  Dynamic holographic endoscopy--ex vivo investigations of malignant tumors in the human stomach.

Authors:  Wolfgang Avenhaus; Björn Kemper; Sabine Knoche; Dirk Domagk; Christopher Poremba; Gert von Bally; Wolfram Domschke
Journal:  Lasers Med Sci       Date:  2005-02-22       Impact factor: 3.161

Review 3.  Small Diameter Xenogeneic Extracellular Matrix Scaffolds for Vascular Applications.

Authors:  Manuela Lopera Higuita; Leigh G Griffiths
Journal:  Tissue Eng Part B Rev       Date:  2019-11-27       Impact factor: 6.389

4.  Fluid-structure interaction in aortic cross-clamping: implications for vessel injury.

Authors:  Henry Y Chen; Jose A Navia; Shoaib Shafique; Ghassan S Kassab
Journal:  J Biomech       Date:  2009-11-01       Impact factor: 2.712

5.  Calcification and identification of metalloproteinases in bovine pericardium after subcutaneous implantation in rats.

Authors:  E Jorge-Herrero; J Turnay; P Calero; N Olmo; I López De Silanes; M Martín Maestro; M A Lizarbe; J L Castillo-Olivares
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

6.  Screening biomaterials with a new in vitro method for potential calcification: porcine aortic valves and bovine pericardium.

Authors:  D Mavrilas; J Kapolos; P G Koutsoukos; D Dougenis
Journal:  J Mater Sci Mater Med       Date:  2004-06       Impact factor: 3.896

7.  Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement.

Authors:  Guangyu Zhu; Munirah Binte Ismail; Masakazu Nakao; Qi Yuan; Joon Hock Yeo
Journal:  PLoS One       Date:  2019-01-30       Impact factor: 3.240

8.  Tissue engineered bovine saphenous vein extracellular matrix scaffolds produced via antigen removal achieve high in vivo patency rates.

Authors:  Manuela Lopera Higuita; Juan F Lopera Giraldo; Tiffany L Sarrafian; Leigh G Griffiths
Journal:  Acta Biomater       Date:  2021-06-27       Impact factor: 10.633

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.