Assessment of cold welding properties of the internal conical interface of two commercially available implant systems.

M R Norton1.   

Abstract

STATEMENT OF PROBLEM: The cone-screw abutment has been shown to diminish micromovement by reducing the burden of component loosening and fracture. However, anecdotal concern for cold welding of cone-screw joints in implant design has been identified as a potential source for lack of retrievability.
PURPOSE: This comparative study evaluated the loosening torque, as a percentage of tightening torque, for the ITI Straumann and Astra Tech (3.5 and 4.0 mm diameters) implant systems, which use an 8-degree and 11-degree internal cone, respectively.
MATERIAL AND METHODS: Implants and abutments from each system were mounted in a torque device, and a range of tightening torques was applied. Loosening torques were then measured, and the influence of conus angle, interfacial surface area, saliva contamination, and time delay to loosening were all assessed.
RESULTS: The loosening torque only exceeded tightening torque at the highest levels, just before component failure, when plastic deformation was expected. For all clinically relevant levels of torque, both in a dry environment and with components bathed in artificial saliva at 37 degrees C, loosening torque was always seen to be 80% to 90% of tightening torque, demonstrating that cold welding does not occur. There was a high correlation between loosening and tightening torque for all systems tested, but no statistical difference when comparing wet versus dry or comparing individual data for each system.
CONCLUSIONS: It can be concluded that for clinically relevant levels of tightening torque, no problems are anticipated with respect to retrievability.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 9922428     DOI: 10.1016/s0022-3913(99)70243-x

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  11 in total

1.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

2.  Cumulative survival rate of Astra Tech implants: a retrospective analysis.

Authors:  Jung-Soo Kim; Joo-Yeon Sohn; Jung-Chul Park; Ui-Won Jung; Chang-Sung Kim; Jae-Hoon Lee; June-Sung Shim; Keun-Woo Lee; Seong-Ho Choi
Journal:  J Periodontal Implant Sci       Date:  2011-04-29       Impact factor: 2.614

3.  A three-dimensional finite element analysis of a passive and friction fit implant abutment interface and the influence of occlusal table dimension on the stress distribution pattern on the implant and surrounding bone.

Authors:  Hasan Sarfaraz; Anoopa Paulose; K Kamalakanth Shenoy; Akhter Hussain
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

4.  Association between dental implants in the posterior region and traumatic occlusion in the adjacent premolars: a long-term follow-up clinical and radiographic analysis.

Authors:  Jae-Hong Lee; Helen Hye-In Kweon; Seong-Ho Choi; Young-Taek Kim
Journal:  J Periodontal Implant Sci       Date:  2016-12-26       Impact factor: 2.614

5.  Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.

Authors:  Saied Nokar; Habib Hajimiragha; Leyla Sadighpour; Azam Sadat Mostafavi
Journal:  Dent Res J (Isfahan)       Date:  2020-12-10

6.  Biomechanical analysis of rigid and non-rigid connection with implant abutment designs for tooth-implant supported prosthesis: A finite element analysis.

Authors:  Yen-Chang Huang; Shinn-Jyh Ding; Cadmus Yuan; Min Yan
Journal:  J Dent Sci       Date:  2021-08-12       Impact factor: 2.080

7.  Influence of the implant abutment types and the dynamic loading on initial screw loosening.

Authors:  Eun-Sook Kim; Soo-Yeon Shin
Journal:  J Adv Prosthodont       Date:  2013-02-28       Impact factor: 1.904

8.  Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates.

Authors:  Christopher Ioannou; Matthew Knight; Luca Daniele; Lee Flueckiger; Ezekiel S L Tan
Journal:  J Orthop Surg Res       Date:  2016-10-17       Impact factor: 2.359

9.  A novel technique to retrieve cold-welded implant abutment: A case series.

Authors:  Mahnaz Arshad; Hossein Ali Mahgoli; Kamran Rasouli; Sina Refoua
Journal:  Clin Case Rep       Date:  2019-08-16

10.  Mechanical Assessment of Fatigue Characteristics between Single- and Multi-Directional Cyclic Loading Modes on a Dental Implant System.

Authors:  Won Hyeon Kim; Eun Sung Song; Kyung Won Ju; Dohyung Lim; Dong-Wook Han; Tae-Gon Jung; Yong-Hoon Jeong; Jong-Ho Lee; Bongju Kim
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.