Literature DB >> 9894940

Comparison of corneal endothelial image analysis by Konan SP8000 noncontact and Bio-Optics Bambi systems.

B A Benetz1, E Diaconu, S J Bowlin, S S Oak, R A Laing, J H Lass.   

Abstract

PURPOSE: Compare corneal endothelial image analysis by Konan SP8000 and Bio-Optics Bambi image-analysis systems.
METHODS: Corneal endothelial images from 98 individuals (191 eyes), ranging in age from 4 to 87 years, with a normal slit-lamp examination and no history of ocular trauma, intraocular surgery, or intraocular inflammation were obtained by the Konan SP8000 noncontact specular microscope. One observer analyzed these images by using the Konan system and a second observer by using the Bio-Optics Bambi system. Three methods of analyses were used: a fixed-frame method to obtain cell density (for both Konan and Bio-Optics Bambi) and a "dot" (Konan) or "corners" (Bio-Optics Bambi) method to determine morphometric parameters.
RESULTS: The cell density determined by the Konan fixed-frame method was significantly higher (157 cells/mm2) than the Bio-Optics Bambi fixed-frame method determination (p<0.0001). However, the difference in cell density, although still statistically significant, was smaller and reversed comparing the Konan fixed-frame method with both Konan dot and Bio-Optics Bambi comers method (-74 cells/mm2, p<0.0001; -55 cells/mm2, p<0.0001, respectively). Small but statistically significant morphometric analyses differences between Konan and Bio-Optics Bambi were seen: cell density, +19 cells/mm2 (p = 0.03); cell area, -3.0 microm2 (p = 0.008); and coefficient of variation, +1.0 (p = 0.003). There was no statistically significant difference between these two methods in the percentage of six-sided cells detected (p = 0.55).
CONCLUSION: Cell densities measured by the Konan fixed-frame method were comparable with Konan and Bio-Optics Bambi's morphometric analysis, but not with the Bio-Optics Bambi fixed-frame method. The two morphometric analyses were comparable with minimal or no differences for the parameters that were studied. The Konan SP8000 endothelial image-analysis system may be useful for large-scale clinical trials determining cell loss; its noncontact system has many clinical benefits (including patient comfort, safety, ease of use, and short procedure time) and provides reliable cell-density calculations.

Entities:  

Mesh:

Year:  1999        PMID: 9894940

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  8 in total

1.  Automated tri-image analysis of stored corneal endothelium.

Authors:  P Gain; G Thuret; L Kodjikian; Y Gavet; P H Turc; C Theillere; S Acquart; J C Le Petit; J Maugery; L Campos
Journal:  Br J Ophthalmol       Date:  2002-07       Impact factor: 4.638

2.  Assessment of a variable frame (polygonal) method to estimate corneal endothelial cell counts after corneal transplantation.

Authors:  S Jonuscheit; M J Doughty; K Ramaesh
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

3.  Corneal endothelial cell density and morphology in low and moderate myopic Chinese eyes.

Authors:  Samaneh Delshad; Jane Mei Chun
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

Review 4.  Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions.

Authors:  Bernard E McCarey; Henry F Edelhauser; Michael J Lynn
Journal:  Cornea       Date:  2008-01       Impact factor: 2.651

5.  Automatic estimation of endothelium cell density in donor corneas by means of fourier analysis.

Authors:  M Foracchia; A Ruggeri
Journal:  Med Biol Eng Comput       Date:  2004-09       Impact factor: 2.602

6.  Specular microscopy ancillary study methods for donor endothelial cell density determination of Cornea Donor Study images.

Authors:  Beth Ann Benetz; Robin L Gal; Katrina J Ruedy; Carmella Rice; Roy W Beck; Andrea D Kalajian; Jonathan H Lass
Journal:  Curr Eye Res       Date:  2006-04       Impact factor: 2.424

7.  Evaluation of corneal endothelial cell damage after vitreoretinal surgery: comparison of different endotamponades.

Authors:  E Cinar; M O Zengin; C Kucukerdonmez
Journal:  Eye (Lond)       Date:  2015-03-13       Impact factor: 3.775

8.  Comparison of manual & automated analysis methods for corneal endothelial cell density measurements by specular microscopy.

Authors:  Jianyan Huang; Jyotsna Maram; Tudor C Tepelus; Cristina Modak; Ken Marion; SriniVas R Sadda; Vikas Chopra; Olivia L Lee
Journal:  J Optom       Date:  2017-08-07
  8 in total