Thermal effect on corneal incisions with different phacoemulsification ultrasonic tips.

H Bissen-Miyajima1, S Shimmura, K Tsubota.   

Abstract

PURPOSE: To evaluate the thermal effects of 3 phacoemulsification ultrasound (US) tips (standard, MicroTip, and Mackool) on the corneal incision based on the tip's position.
SETTING: Department of Ophthalmology, Tokyo Dental College, Tokyo, Japan.
METHODS: Thermal changes were recorded with an infrared thermal video during phacoemulsification in 18 porcine eyes. The 3 US tips were placed in 3 positions (primary, roofing, and flanking) in the corneal incision, and the resultant temperature changes were recorded.
RESULTS: With the 3 US tips in the primary position, the temperature remained under 36 degrees C. When the tips were in the roofing position, a marked temperature increase (51.4 degrees C) was observed with the standard tip; with the MicroTip, a temperature increase occurred but was lower than with the standard tip; with the Mackool tip, the temperature remained stable. The temperature around the incision site increased with all tips when they were moved toward the corner of the incision, which resulted in direct contact with the silicone sleeve. This change was most obvious with the standard tip (temperature 51.5 degrees C), and a wound burn was observed. With the MicroTip, the temperature increase was moderate (44.2 degrees C), and the wound burn area was smaller. With the Mackool tip, the temperature increased the least (42.1 degrees C).
CONCLUSION: The position of the US tips resulted in a temperature increase in the incision area and caused wound burn. Newly developed US tips could minimize this complication.

Entities:  

Mesh:

Year:  1999        PMID: 9888078     DOI: 10.1016/s0886-3350(99)80012-1

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Comparison of peristaltic and Venturi pumps in bimanual microincisional cataract surgery.

Authors:  Hande Karaguzel; Aylin Karalezli; Bekir Sitki Aslan
Journal:  Int Ophthalmol       Date:  2008-10-14       Impact factor: 2.031

2.  Heat profiling of phacoemulsification tip using a thermal scanning camera.

Authors:  Wei Kiong Ngo; Louis W Lim; Colin S H Tan; Wee Jin Heng
Journal:  Int Ophthalmol       Date:  2013-03-20       Impact factor: 2.031

3.  Investigation of scleral thermal injuries caused by ultrasonic pars plana phacoemulsification and aspiration using pig eyes.

Authors:  Takaki Sato; Toru Yasuhara; Masanori Fukumoto; Masashi Mimura; Takatoshi Kobayashi; Teruyo Kida; Shota Kojima; Hidehiro Oku; Tsunehiko Ikeda
Journal:  Int Ophthalmol       Date:  2018-10-23       Impact factor: 2.031

4.  Laboratory assessment of thermal characteristics of three phacoemulsification tip designs operated using torsional ultrasound.

Authors:  Jaime Zacharias
Journal:  Clin Ophthalmol       Date:  2016-06-13

5.  In water thermal imaging comparison of the Alcon legacy and AMO sovereign phacoemulsification systems.

Authors:  Kevin M Miller; Michael D Olson
Journal:  Open Ophthalmol J       Date:  2008-02-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.