Literature DB >> 9870021

Recording of pacing stimulus artifacts by endovascular defibrillation lead systems: comparison of true and integrated bipolar circuits.

V Menz1, D Schwartzman, D Drachman, J J Michele, S M Dillon.   

Abstract

BACKGROUND: The occurrence ICD undersensing of ventricular fibrillation due to the presence of a pacing stimulus artifact (PSA) is in part related to the amplitude of the artifact recorded on the ICD rate sensing circuit. There is little comparative data regarding PSA amplitude recorded by commercial ICD rate-sensing circuits.
PURPOSE: To compare PSA amplitude recorded by commercial endovascular defibrillation leads utilizing integrated or true bipolar sensing circuits.
METHODS: Nineteen large (60-120 kg) pigs were utilized. Two different commercial endovascular defibrillation leads were evaluated, each with its distal tip located at the right ventricular apex: (1) Medtronic Transvene; and (2) CPI Endotak. Three different rate-sensing circuits were evaluated: (1) Transvene true bipolar (tip-ring); (2) Transvene integrated bipolar (tip-coil); and (3) Endotak integrated bipolar (tip-coil). Using a separate pacing lead located at the left ventricular apex (n = 19 animals) or right ventricular outflow tract (n = 10 animals), pacing was performed at a pulse width of 0.5 milliseconds at outputs of 1.5, 5 and 10 volts. PSA amplitude was recorded at each output by each circuit.
RESULTS: During pacing from the left ventricular apex, at each pacing output voltage the PSA amplitude recorded by the true bipolar circuit (0.6 +/- 0.1 mV at 1.5 volts, 2.0 +/- 0.5 mV at 5 volts, 3.7 +/- 0.8 mV at 10 volts) was significantly smaller than recorded by the Transvene integrated circuit (1.4 +/- 0.3 mV at 1.5 volts, 3.8 +/- 0.7 mV at 5 volts, 4.1 +/- 0.8 mV at 10 volts) or the Endotak integrated circuit (1.8 +/- 0.4 mV at 1.5 volts, 4.2 +/- 1.0 mV at 5 volts, 6.3 +/- 1.8 mV at 10 volts). During pacing from the right ventricular outflow tract, at each pacing output voltage the PSA amplitude recorded by the true bipolar circuit (0.7 +/- 0.1 mV at 1.5 volts, 1.7 +/- 0.4 mV at 5 volts, 4.0 +/- 0.7 mV at 10 volts) was significantly smaller than recorded by the Transvene integrated circuit (1.1 +/- 0.4 mV at 1.5 volts, 3.9 +/- 1.2 mV at 5 volts, 7.5 +/- 1.8 mV at 10 volts) or the Endotak integrated circuit (1.6 +/- 0.7 mV at 1.5 volts, 4.3 +/- 1.7 mV at 5 volts, 7.5 +/- 2.6 mV at 10 volts). For both pacing sites, the PSA amplitude recorded by the two integrated circuits was not significantly different.
CONCLUSIONS: For a given pacing output voltage, PSA amplitude recorded by commercial endovascular rate sensing/defibrillation leads is greater when the sensing circuit is integrated than when it is true bipolar. These data may be helpful in planning ICD implantation in patients with previously implanted permanent pacemakers.

Entities:  

Mesh:

Year:  1998        PMID: 9870021     DOI: 10.1023/a:1009789004697

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  5 in total

1.  Combined automatic implantable cardioverter-defibrillator and pacemaker systems: implantation techniques and follow-up.

Authors:  A E Epstein; G N Kay; V J Plumb; R B Shepard; J K Kirklin
Journal:  J Am Coll Cardiol       Date:  1989-01       Impact factor: 24.094

Review 2.  ICD-antiarrhythmic drug and ICD-pacemaker interactions.

Authors:  S E Brode; D Schwartzman; D J Callans; C D Gottlieb; F E Marchlinski
Journal:  J Cardiovasc Electrophysiol       Date:  1997-07

3.  Unipolar pacer artifacts induced failure of an automatic implantable cardioverter/defibrillator to detect ventricular fibrillation.

Authors:  S G Kim; S Furman; L E Waspe; R Brodman; J D Fisher
Journal:  Am J Cardiol       Date:  1986-04-01       Impact factor: 2.778

4.  Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators.

Authors:  H Calkins; J Brinker; E P Veltri; T Guarnieri; J H Levine
Journal:  J Am Coll Cardiol       Date:  1990-09       Impact factor: 24.094

5.  Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers.

Authors:  R Brooks; H Garan; B A McGovern; J N Ruskin
Journal:  Am Heart J       Date:  1995-01       Impact factor: 4.749

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.