Literature DB >> 9869026

Monte Carlo calculations of electron beam output factors for a medical linear accelerator.

A Kapur1, C M Ma, E C Mok, D O Findley, A L Boyer.   

Abstract

The purpose of this study was to investigate the application of the Monte Carlo technique to the calculation and analysis of output factors for electron beams used in radiotherapy. The code EGS4/BEAM was used to obtain phase-space files for 6, 12 and 20 MeV clinical electron beams from a scattering-foil linac (Varian Clinac 2100C) for a clinically representative range of applicator and square or rectangular insert combinations. The source-to-surface distance used was 100 cm. The field sizes ranged from 1 x 1 cm2 to 20 x 20 cm2. These phase-space files were analysed to study the intrinsic beam characteristics and used as source input for relative dose and output factor computations in homogeneous water phantoms using the code EGS4/DOSXYZ. The calculated relative central-axis depth-dose and transverse dose profiles at various depths of clinical interest agreed with the corresponding measured dose profiles to within 2% of the maximum dose. Calculated output factors for the fields studied agreed with measured output factors to about 2%. This demonstrated that for the Varian Clinac 2100C linear accelerator, electron beam dose calculations in homogeneous water phantoms can be performed accurately at the 2% level using Monte Carlo simulations.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9869026     DOI: 10.1088/0031-9155/43/12/007

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Monte Carlo simulations of electron beams collimated with a dual electron multileaf collimator: a feasibility study.

Authors:  S O Inyang; A C Chamberlain
Journal:  Radiol Phys Technol       Date:  2009-07-04

2.  Monte Carlo calculations for absolute dosimetry to determine machine outputs for proton therapy fields.

Authors:  Harald Paganetti
Journal:  Phys Med Biol       Date:  2006-05-17       Impact factor: 3.609

Review 3.  Monte Carlo systems used for treatment planning and dose verification.

Authors:  Lorenzo Brualla; Miguel Rodriguez; Antonio M Lallena
Journal:  Strahlenther Onkol       Date:  2016-11-25       Impact factor: 3.621

4.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

Review 5.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

6.  Design and evaluation of electron beam energy degraders for breast boost irradiation.

Authors:  Jong In Park; Sung Whan Ha; Jung-In Kim; Hyunseok Lee; Jaegi Lee; Il Han Kim; Sung-Joon Ye
Journal:  Radiat Oncol       Date:  2016-08-31       Impact factor: 3.481

7.  Monte Carlo calculations of output factors for clinically shaped electron fields.

Authors:  Julius V Turian; Brett D Smith; Damian A Bernard; Katherine L Griem; James C Chu
Journal:  J Appl Clin Med Phys       Date:  2004-04-01       Impact factor: 2.102

8.  Electron beam therapy at extended source-to-surface distance: a Monte Carlo investigation.

Authors:  Tuathan P O'Shea; Mark J Foley; David Rajasekar; Patrick A Downes; Wil van der Putten; Margaret Moore; Andrew Shearer
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

9.  Convolution-based modified Clarkson integration (CMCI) for electron cutout factor calculation.

Authors:  Jina Chang; Mu-Han Lin; Weiguo Lu; Mingli Chen; Steve Jiang
Journal:  J Appl Clin Med Phys       Date:  2018-02-03       Impact factor: 2.102

10.  Benchmarking of Siemens Linac in Electron Modes: 8-14 MeV Electron Beams.

Authors:  H Dowlatabadi; A A Mowlavi; M Ghorbani; S Mohammadi; F Akbari
Journal:  J Biomed Phys Eng       Date:  2018-06-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.