Magnification characteristics of fundus imaging systems.

A R Rudnicka1, R O Burk, D F Edgar, F W Fitzke.   

Abstract

OBJECTIVE: To compare the magnification properties of 11 different fundus cameras (including 1 stereo fundus camera), a Rodenstock (infrared) scanning laser ophthalmoscope (SLO), the Heidelberg Laser Tomographic Scanner (LTS), and the Heidelberg Retina Tomograph (HRT).
DESIGN: A cross-sectional study of the relationship between the true size of a fundus feature and its photographic-computer image in 14 different fundus imaging devices. This relationship was evaluated for each instrument using a model eye adjusted for axial ametropia between +11 diopter (D) and -14 D. To simulate refractive ametropia, the "crystalline lens" was removed to render the model eye aphakic, and the axial length was adjusted to give aphakic ametropia from emmetropia to +20 D. MAIN OUTCOME MEASURES: A correction factor (p) was calculated for each instrument, which can be used in calculations for determining true retinal size.
RESULTS: The following were found to be of telecentric construction, Zeiss Oberkochen (WS240 Heidelberg), Zeiss Oberkochen (UK), Zeiss Oberkochen (Cologne), Nikon NF505, Kowa RCXV, SLO prototype (UK), LTS, and the HRT, and each exhibited a constant relationship between p and degree of ametropia of the model eye. The Canon CF6OU, Canon CF6OS, Canon CR4-45NM, Nidek 3-DX, Olympus GRCW, and Carl Zeiss Jena Retinophot were found not to be telecentric and exhibited a linear relationship between p and degree of ametropia of the model eye. For all instruments, p remained unchanged for axial and refractive ametropias of the same degree.
CONCLUSIONS: The study has shown that not all fundus imaging systems are telecentric, so the use of a single magnification correction value may not be appropriate. These findings have important implications for the way in which true retinal size calculations are performed. Examples are given to show how the tabulated values of correction factors can be used for both telecentric and nontelecentric cameras in image size calculations.

Mesh:

Year:  1998        PMID: 9855145     DOI: 10.1016/S0161-6420(98)91214-3

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  14 in total

1.  Digital analysis of choroidal neovascularisation in consecutive fluorescein angiograms for use in longitudinal clinical trials.

Authors:  C Bellmann; D W Miller; K Mehltretter; F Schütt; J Jorzik; K Unnebrink; F G Holz
Journal:  Br J Ophthalmol       Date:  2003-07       Impact factor: 4.638

2.  Magnification-corrected photodynamic therapy.

Authors:  Siamak Ansari-Shahrezaei; Erdem Ergun; Robert Chong; Adnan Tufail; Andreas Wedrich; Michael Stur
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-10-06       Impact factor: 3.117

3.  The effects of ocular magnification on Spectralis spectral domain optical coherence tomography scan length.

Authors:  Irene Ctori; Stephen Gruppetta; Byki Huntjens
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-01-10       Impact factor: 3.117

4.  The influence of axial length on confocal scanning laser ophthalmoscopy and spectral-domain optical coherence tomography size measurements: a pilot study.

Authors:  T Röck; B Wilhelm; K U Bartz-Schmidt; D Röck
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-02-23       Impact factor: 3.117

5.  A portable, contact animal fundus imaging system based on Rol's GRIN lenses.

Authors:  Victor Hernandez; Thomas Albini; William Lee; Cornelis Rowaan; Derek Nankivil; Esdras Arrieta; Jean-Marie Parel
Journal:  Vet Ophthalmol       Date:  2011-08-08       Impact factor: 1.644

Review 6.  Lessons learned from quantitative fundus autofluorescence.

Authors:  Janet R Sparrow; Tobias Duncker; Kaspar Schuerch; Maarjaliis Paavo; Jose Ronaldo Lima de Carvalho
Journal:  Prog Retin Eye Res       Date:  2019-08-28       Impact factor: 21.198

7.  A clinical planning module for adaptive optics SLO imaging.

Authors:  Gang Huang; Xiaofeng Qi; Toco Y P Chui; Zhangyi Zhong; Stephen A Burns
Journal:  Optom Vis Sci       Date:  2012-05       Impact factor: 1.973

8.  Assessment of Macular Microvasculature in Healthy Eyes of Infants and Children Using OCT Angiography.

Authors:  S Tammy Hsu; Hoan T Ngo; Sandra S Stinnett; Nathan L Cheung; Robert J House; Michael P Kelly; Xi Chen; Laura B Enyedi; S Grace Prakalapakorn; Miguel A Materin; Mays A El-Dairi; Glenn J Jaffe; Sharon F Freedman; Cynthia A Toth; Lejla Vajzovic
Journal:  Ophthalmology       Date:  2019-07-15       Impact factor: 12.079

9.  Magnification-corrected indirect biomicroscopy of the optic nerve head.

Authors:  Siamak Ansari-Shahrezaei; Michael Stur
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-09-29       Impact factor: 3.117

10.  Comparison of glaucoma diagnostic accuracy of macular ganglion cell complex thickness based on nonhighly myopic and highly myopic normative database.

Authors:  Henry Shen-Lih Chen; Chun-Hsiu Liu; Da-Wen Lu
Journal:  Taiwan J Ophthalmol       Date:  2016-02-13
View more