Evaluation of a long-range transmitter for use with a magnetic tracking device in motion analysis.

J S Day1, G A Dumas, D J Murdoch.   

Abstract

The suitability of a long-range transmitter was evaluated for use with a Polhemus Fastrak magnetic tracking device in kinesiologic studies. Performance was judged by comparing positional and rotational accuracy to a standard transmitter. Data were obtained at distances of up to 2.7 and 5.0 m for the standard and long-range transmitters, respectively. Use of the long-range transmitter improved system performance allowing reproducible measurements at a greater distance. However, it is necessary to calibrate the system in each new test environment as there can be significant distortion of the magnetic field.

Mesh:

Year:  1998        PMID: 9840763     DOI: 10.1016/s0021-9290(98)00089-x

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712