Complete calibration of a stereo photogrammetric system through control points of unknown coordinates.

P Cerveri1, N A Borghese, A Pedotti.   

Abstract

This paper presents a new method for calibrating a video 3D stereo-photogrammetric system. The external parameters and the focal lengths of the cameras are determined from the epipolar constraint and the principal points are computed through the minimisation of a cost function carried out through an evolutionary optimisation. The method has been made more robust with a deterministic annealing procedure of the search region amplitude. Calibration is carried out by moving a rigid bar, carrying two markers on its extremities, inside the working volume. The distance between the two markers is the only measure required. Tests on real data are reported which show that the obtained accuracy is comparable to the one achieved calibrating with control points of known 3D coordinates.

Mesh:

Year:  1998        PMID: 9840759     DOI: 10.1016/s0021-9290(98)00104-3

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  3 in total

1.  A fast method for calibrating video-based motion analysers using only a rigid bar.

Authors:  N A Borghese; P Cerveri; P Rigiroli
Journal:  Med Biol Eng Comput       Date:  2001-01       Impact factor: 2.602

2.  Normative 3D opto-electronic stereo-photogrammetric posture and spine morphology data in young healthy adult population.

Authors:  Moreno D'Amico; Edyta Kinel; Piero Roncoletta
Journal:  PLoS One       Date:  2017-06-22       Impact factor: 3.240

3.  Action Sport Cameras as an Instrument to Perform a 3D Underwater Motion Analysis.

Authors:  Gustavo R D Bernardina; Pietro Cerveri; Ricardo M L Barros; João C B Marins; Amanda P Silvatti
Journal:  PLoS One       Date:  2016-08-11       Impact factor: 3.240