The efficacy of the high volume counterpulsation technique at very low levels of aortic pressure.

C E Charitos1, J N Nanas, D A Kontoyiannis, S N Nanas, G Z Stamatopoulos, A C Rapti, S F Stamatelopoulos, S D Moulopoulos.   

Abstract

BACKGROUND: The objective of the present study was to identify the hemodynamic variables, that preclude the effectiveness of the counterpulsation technique.
METHODS: In order to achieve high counterpulsation volume two devices were used simultaneously. The para-aortic counterpulsation device pumping 60-70 ml was implanted on the ascending aorta, and the intra-aortic balloon (20 ml) was placed in the descending aorta of 14 dogs with severe cardiogenic shock. Both devices were synchronized to provide aortic diastolic augmentation. The mechanical assistance provided by these devices was considered adequate when the mean aortic pressure during the supportive period was at least 50 mmHg, a pressure compatible with life.
RESULTS: The mechanical assistance provided by the two devices was adequate in none (0%) of the cases with a systolic aortic pressure below 30 mmHg (Group I), in 71% of the cases with a systolic aortic pressure between 30-56 mmHg (Group II), and in 100% of the cases with a systolic aortic pressure above 56 mmHg (Group III). In group II the left ventricular end-diastolic pressure was higher in the cases responding adequately to mechanical assistance (Subgroup IIa) than in the ones where no adequate response was obtained (Subgroup IIb).
CONCLUSIONS: The very low levels of the systolic aortic pressure was the most important factor precluding the adequate effectiveness of the counterpulsation technique.

Entities:  

Mesh:

Year:  1998        PMID: 9833723

Source DB:  PubMed          Journal:  J Cardiovasc Surg (Torino)        ISSN: 0021-9509            Impact factor:   1.888


  4 in total

1.  Superior performance of a paraaortic counterpulsation device compared to the intraaortic balloon pump.

Authors:  John V Terrovitis; Christos E Charitos; Elias J Tsolakis; Paraskevi Dolou; Charalampos N Pierrakos; Kostas X Siafakas; John N Nanas
Journal:  World J Surg       Date:  2003-11-05       Impact factor: 3.352

Review 2.  Reverse cardiac remodeling enabled by mechanical unloading of the left ventricle.

Authors:  Konstantinos G Malliaras; John V Terrovitis; Stavros G Drakos; John N Nanas
Journal:  J Cardiovasc Transl Res       Date:  2008-09-30       Impact factor: 4.132

3.  A novel technique for intra-aortic balloon positioning in the intensive care unit.

Authors:  Vijish Venugopal
Journal:  J Extra Corpor Technol       Date:  2012-09

4.  Comparison of effects of extra-thoracic paraaortic counterpulsation to intraaortic balloon pump on circulatory support in acute heart failure.

Authors:  Jie-Min Zhang; Xiao-Cheng Liu; Zhi-Gang Liu; Long Zhao; Li Yang; Tian-Wen Liu; Guo-Wei He
Journal:  J Cardiothorac Surg       Date:  2015-11-24       Impact factor: 1.637

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.