[Evaluation of the Humphrey perimetry programs SITA Standard and SITA Fast in normal probands and patients with glaucoma].

J P Nordmann1, F Brion, P Hamard, D Mouton-Chopin.   

Abstract

PURPOSE: To evaluate the new threshold strategies of the Humphrey Field Analyser, SITA Standard and SITA Fast, in normals and glaucoma patients.
METHODS: Fifty healthy volunteers and 50 glaucoma patients were tested twice with three algorithms: Full Threshold, SITA Standard and SITA Fast in two sessions. The second test was taken into account to eliminate learning effect. Glaucoma patients were chosen to cover a large range of deficits. Quantitative analysis of global indexes was performed as well as qualitative comparison of visual fields by a trained optometrist.
RESULTS: Compared to Full Threshold, testing time was reduced by 51% in normals and 49% in glaucoma with SITA Standard and by 72% and 69% with SITA Fast, respectively. In glaucoma, mean testing time dropped from 16'01" with Full Threshold to 8'05" with SITA Standard and to 4'55" with SITA Fast. Quantitative comparison showed a high correlation between indexes (MD Full Threshold vs MD SITA Standard: r = 0.98, p < 0.01; PSD Full Threshold vs PSD SITA Standard: r = 0.94, p < 0.01). However SITA Standard and SITA Fast had a tendency to underestimate visual field defects. In glaucomatous subjects, mean MD ameliorated by 1.01 dB with SITA Standard and by 1.71 dB with SITA Fast. Mean PSD ameliorated by 0.04 dB and 0.43 dB, respectively. Qualitative analysis confirmed these last results showing artificial slight or important improvement of visual fields in 21% of the cases with SITA Standard and 30% with SITA Fast, compared to Full Threshold.
CONCLUSION: SITA Standard and SITA Fast strategies improve dramatically visual field testing time, by 51% and 72% respectively. However, one should be cautious in comparing results obtained with different algorithms, as these new strategies improve the mean defect and, to a lesser extent, the localized defect.

Entities:  

Mesh:

Year:  1998        PMID: 9833219

Source DB:  PubMed          Journal:  J Fr Ophtalmol        ISSN: 0181-5512            Impact factor:   0.818


  5 in total

1.  Can Visual Field Progression be Predicted by Confocal Scanning Laser Ophthalmoscopic Imaging of the Optic Nerve Head in Glaucoma? (An American Ophthalmological Society Thesis).

Authors:  John Danias; Janet Serle
Journal:  Trans Am Ophthalmol Soc       Date:  2015

2.  The Effect of Transitioning from SITA Standard to SITA Faster on Visual Field Performance.

Authors:  Alex T Pham; Pradeep Y Ramulu; Michael V Boland; Jithin Yohannan
Journal:  Ophthalmology       Date:  2021-03-30       Impact factor: 14.277

3.  Comparison of 30-2 Standard and Fast programs of Swedish Interactive Threshold Algorithm of Humphrey Field Analyzer for perimetry in patients with intracranial tumors.

Authors:  Manav Deep Singh; Kanika Jain
Journal:  Indian J Ophthalmol       Date:  2017-11       Impact factor: 1.848

4.  Differences in visual field loss pattern when transitioning from SITA standard to SITA faster.

Authors:  Christopher T Le; Jacob Fiksel; Pradeep Ramulu; Jithin Yohannan
Journal:  Sci Rep       Date:  2022-04-29       Impact factor: 4.996

Review 5.  Homonymous hemianopia: challenges and solutions.

Authors:  Denise Goodwin
Journal:  Clin Ophthalmol       Date:  2014-09-22