Aortic function in patients during intra-aortic balloon pumping determined by the pressure-diameter relation.

C Stefanadis1, J Dernellis, E Tsiamis, C Stratos, I Kallikazaros, P Toutouzas.   

Abstract

OBJECTIVE: The physiologic basis for the hemodynamic and clinical improvement achieved by the use of intra-aortic balloon pumping in patients with cardiogenic shock has not been clarified in all its aspects. This study evaluated the possible contribution of pump-induced alterations of aortic mechanics to the overall benefit gained by the implementation of this therapeutic modality in patients with acute heart failure of ischemic origin.
METHODS: The aortic pressure-diameter relation was obtained by use of an intravascular catheter for aortic diameter measurements developed in our institution and previously validated, simultaneously with a catheter-tip micromanometer for aortic pressure measurements at the same aortic level. Aortic function indices were compared before and during intra-aortic balloon pumping in 12 patients with cardiogenic shock.
RESULTS: Intra-aortic balloon pumping increased cardiac index and aortic distensibility by 24% and 30%, respectively, and reduced myocardial oxygen demand by 31% (P <.001 for all alterations). Energy loss caused by aortic wall viscosity increased by 207% (P <. 001). The aortic diameter augmentation index increased by 68% (P <. 001); the aortic pressure augmentation index decreased by 117% (P <. 001). Linear regression analysis showed that cardiac index and myocardial oxygen demand were related with the aortic stiffness constant both before and during intra-aortic balloon pumping.
CONCLUSIONS: During intra-aortic balloon pumping, aortic distensibility was improved, and wave reflection from the arterial periphery was reduced. The relationship between cardiac index and myocardial oxygen demand and aortic stiffness suggests that improvement of the elastic properties of the aorta was an important mechanism by which intra-aortic balloon pumping improved circulatory function.

Entities:  

Mesh:

Year:  1998        PMID: 9832698     DOI: 10.1016/S0022-5223(98)70058-3

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

1.  Intra-aortic balloon pumping reduces the increased arterial load caused by acute cardiac depression, modifying central and peripheral load determinants in a time- and flow-related way.

Authors:  Daniel Bia; Edmundo I Cabrera-Fischer; Yanina Zócalo; Ricardo L Armentano
Journal:  Heart Vessels       Date:  2011-11-09       Impact factor: 2.037

2.  Monitoring of arterial stiffness indices by applanation tonometry and pulse wave analysis: reproducibility at low blood pressures.

Authors:  Theodoros G Papaioannou; Kimon S Stamatelopoulos; Elias Gialafos; Charalambos Vlachopoulos; Emmanouil Karatzis; John Nanas; John Lekakis
Journal:  J Clin Monit Comput       Date:  2004-04       Impact factor: 2.502

Review 3.  Evidence of clinical efficacy of counterpulsation therapy methods.

Authors:  M Capoccia; C T Bowles; J R Pepper; N R Banner; A R Simon
Journal:  Heart Fail Rev       Date:  2015-05       Impact factor: 4.214