Effect of flow rate and insulin priming on the recovery of insulin from microbore infusion tubing.

M Fuloria1, M A Friedberg, R H DuRant, J L Aschner.   

Abstract

BACKGROUND: A retrospective medical record review of 13 consecutive, hyperglycemic, extremely low birth weight (ELBW) infants treated with continuous insulin infusions revealed a 14- to 24-hour delay (mean, 19 hours) in blood glucose normalization despite stepwise increases in insulin infusion rates.
OBJECTIVE: This in vitro study examined the effects of flow rate and insulin priming on insulin recovery from polyvinyl chloride (PVC) tubing and polyethylene (PE)-lined PVC tubing infused with a standard insulin stock solution.
METHODS: Stock insulin solution (0.2 U/mL) was infused through microbore PVC or PE-lined tubing at flow rates of 0.05 and 0.2 mL/h. To determine if saturation of nonspecific binding sites would alter effluent insulin concentration, we compared insulin recovery from tubing previously flushed with the stock solution and tubing primed with 5 U/mL of insulin for 20 minutes. Effluent samples, which were collected at baseline and at six time points during a 24-hour period, were immediately frozen at -20 degreesC. Insulin concentration was measured by IMx immunoassay. Data were analyzed using general linear modeling with repeated measures.
RESULTS: At 0.05 mL/h flow rate, insulin recovery from unprimed PVC tubing at 1, 2, 4, and 8 hours was 17%, 11%, 27%, and 55%, respectively, with 100% recovery at 24 hours. From insulin-primed tubing, insulin recovery was approximately 70% at 1, 2, and 4 hours, and close to 100% at 8 hours. At a faster flow rate of 0.2 mL/h, insulin recovery at 1, 2, 4, and 8 hours was 22%, 38%, 67%, and 75% vs 42%, 85%, 91% and 95% from unprimed and insulin-primed PVC tubing, respectively. Similar results were obtained from unprimed and insulin-primed PE-lined tubing at 0.2 mL/h flow rate.
CONCLUSIONS: Priming of microbore tubing with 5 U/mL of insulin solution for 20 minutes to block nonspecific binding sites enhances delivery of a standard insulin stock at infusion rates typically used to treat hyperglycemic ELBW infants. We conclude that priming the tubing with a higher concentration of insulin before initiation of standard insulin infusion therapy should accelerate achievement of steady-state insulin delivery and correction of hyperglycemia in ELBW infants.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9832576     DOI: 10.1542/peds.102.6.1401

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  11 in total

1.  Clinical Recommendations for Managing the Impact of Insulin Adsorptive Loss in Hospital and Diabetes Care.

Authors:  Jennifer L Knopp; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2020-04-24

2.  The impact of insulin adsorption onto the infusion sets in the adult intensive care unit.

Authors:  Therese Jakobsson; Robert Shulman; Hardyal Gill; Kevin Taylor
Journal:  J Diabetes Sci Technol       Date:  2009-01

3.  Capacity of Infusion Lines for Insulin Adsorption: Effect of Flow Rate on Total Adsorption.

Authors:  Jennifer L Knopp; Kaia Bishop; Theodore Lerios; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2019-09-27

4.  Effect of Tubing Flush or Preconditioning on Available Insulin Concentration for IV Infusion: A Pilot Project.

Authors:  Brandi Newby; Daniel T Holmes
Journal:  Can J Hosp Pharm       Date:  2017-08-31

5.  Blood glucose controller for neonatal intensive care: virtual trials development and first clinical trials.

Authors:  Aaron Le Compte; J Geoffrey Chase; Adrienne Lynn; Chris Hann; Geoffrey Shaw; Xing-Wei Wong; Jessica Lin
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

6.  The effect of insulin infusion upon protein metabolism in neonates on extracorporeal life support.

Authors:  Michael S D Agus; Patrick J Javid; Hannah G Piper; David Wypij; Christopher P Duggan; Daniel P Ryan; Tom Jaksic
Journal:  Ann Surg       Date:  2006-10       Impact factor: 12.969

7.  The effect of tubing dwell time on insulin adsorption during intravenous insulin infusions.

Authors:  Cecilia D Thompson; Jessica Vital-Carona; E Vincent S Faustino
Journal:  Diabetes Technol Ther       Date:  2012-07-02       Impact factor: 6.118

8.  Intravenous insulin decreases protein breakdown in infants on extracorporeal membrane oxygenation.

Authors:  Michael S D Agus; Patrick J Javid; Daniel P Ryan; Tom Jaksic
Journal:  J Pediatr Surg       Date:  2004-06       Impact factor: 2.545

9.  In vitro assessment of the influence of intravenous extension set materials on insulin aspart drug delivery.

Authors:  Morgane Masse; Mickael Maton; Stéphanie Genay; Nicolas Blanchemain; Christine Barthélémy; Bertrand Décaudin; Pascal Odou
Journal:  PLoS One       Date:  2018-08-16       Impact factor: 3.240

10.  Perioperative glycaemic control for preterm infant with transient neonatal hyperglycaemia and gastroschisis.

Authors:  Sirirat Rattana-Arpa; Saowaphak Lapmahapaisan; Arunotai Siriussawakul
Journal:  BMC Res Notes       Date:  2016-03-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.