The design of a vibration transducer to monitor the integrity of dental implants.

P Cawley1, B Pavlakovic, D N Alleyne, R George, T Back, N Meredith.   

Abstract

Bone-anchored titanium implants are being used increasingly to provide support for prostheses replacing missing teeth in edentulous and partially dentate patients. A technique is required to monitor bone formation at the implant-tissue interface during healing, and also to check whether there has been bone loss from around the top of the implant. One possible method is to screw a beam into the implanted fixture and to measure the first flexural resonance frequency of the resulting system. This resonance frequency is affected by both the exposed length of fixture and the stiffness of the interface between the implant and the bone. This paper describes the design of a beam-like transducer for clinical trials of the technique. The sensitivity of the transducer resonance frequency to the changes of interest is dependent on the thickness and length of the beam element. However, the choice of these dimensions is constrained by the need to avoid closely spaced resonances. The performance of different transducer shapes and the influence of the thickness and length of the beam element in the transducer has been studied. The results have been used to finalize a transducer design for the clinical trials.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9769694     DOI: 10.1243/0954411981534042

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  4 in total

1.  Effects on the torsional vibration behavior in the investigation of dental implant osseointegration using resonance frequency analysis: a numerical approach.

Authors:  Min Zhai; Bing Li; Dehua Li
Journal:  Med Biol Eng Comput       Date:  2017-02-07       Impact factor: 2.602

2.  Customized SmartPeg for measurement of resonance frequency of mini dental implants.

Authors:  Jagjit Singh Dhaliwal; Rubens F Albuquerque; Ali Fakhry; Sukhbir Kaur; Jocelyne S Feine
Journal:  Int J Implant Dent       Date:  2017-02-01

3.  Comparative evaluation of the stability of two different dental implant designs and surgical protocols-a pilot study.

Authors:  David E Simmons; Pooja Maney; Austin G Teitelbaum; Susan Billiot; Lomesh J Popat; A Archontia Palaiologou
Journal:  Int J Implant Dent       Date:  2017-05-02

4.  Mechanical Properties of the Periodontal System and of Dental Constructs Deduced from the Free Response of the Tooth.

Authors:  Cosmin Sinescu; Virgil-Florin Duma; Dorin Dodenciu; Stefan Stratul; Marius Manole; Gheorghe Eugen Draganescu
Journal:  J Healthc Eng       Date:  2018-09-17       Impact factor: 2.682

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.