Axial, instantaneous, and refractive formulas in computerized videokeratography of normal corneas.

S Bafna1, T Kohnen, D D Koch.   

Abstract

PURPOSE: To compare the values for corneal power determined by the axial, instantaneous and refractive formulas when imaging normal human corneas using computerized videokeratography.
SETTING: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
METHODS: This prospective clinical trial involved 60 corneas of 30 normal volunteers. Computerized videokeratography was performed to determine corneal power at the center and the 1, 3, 5, and 7 mm zones using the 3 formulas.
RESULTS: Mean central corneal power was 42.86 diopters (D) with each of the formulas. The mean corneal powers for the axial, instantaneous, and refractive formulas were 43.09, 43.21, and 42.98 D at the 1 mm zone; 43.10, 42.92, and 43.46 D at the 3 mm zone; 42.75, 41.63, and 44.02 at the 5 mm zone; 42.21, 40.30, and 44.79 D at the 7 mm zone, respectively. The differences among powers for the 3 formulas at the 3, 5, and 7 mm zones were statistically significant (P < .01).
CONCLUSION: In normal corneas, clinically significant differences exist in the corneal power values calculated by the axial, instantaneous, and refractive formulas.

Entities:  

Mesh:

Year:  1998        PMID: 9768390     DOI: 10.1016/s0886-3350(98)80009-6

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Transient changes in refractive error and corneal tomography after 24-h continuous monitoring of intraocular pressure patterns with a contact lens sensor.

Authors:  Atsuya Miki; Miho Kumoi; Naoyuki Maeda; Shizuka Koh; Kenji Matsushita; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2020-02-13       Impact factor: 2.447

2.  Keratoconus: classification scheme based on videokeratography and clinical signs.

Authors:  Xiaohui Li; Huiying Yang; Yaron S Rabinowitz
Journal:  J Cataract Refract Surg       Date:  2009-09       Impact factor: 3.351

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.