Literature DB >> 9735896

Application of reconstruction-based scatter compensation to thallium-201 SPECT: implementations for reduced reconstructed image noise.

D J Kadrmas1, E C Frey, B M Tsui.   

Abstract

Scatter compensation in Tl-201 single photon emission computed tomography (SPECT) presents an interesting challenge because of the multiple emission energies and relatively large proportion of scattered photons. In this paper, we present a simulation study investigating reconstructed image noise levels arising from various implementations of iterative reconstruction-based scatter compensation (RBSC) in Tl-201 SPECT. A two-stage analysis was used to study single and multiple energy window implementations of reconstruction-based scatter compensation, and RBSC was compared to the upper limits on performance for other approaches to handling scatter. In the first stage, singular value decomposition of the system transfer matrix was used to analyze noise levels in a manner independent of the choice of reconstruction algorithm, providing results valid across a wide range of regularizations. In the second stage, the data were reconstructed using maximum-likelihood expectation-maximization, and the noise properties of the resultant images were analyzed. The best RBSC performance was obtained using multiple energy windows, one for each emission photopeak, and RBSC outperformed the upper limit on subtraction-based compensation methods. Implementing RBSC with the correct choice of energy window acquisition scheme is a promising method for performing scatter compensation for Tl-201 SPECT.

Mesh:

Substances:

Year:  1998        PMID: 9735896      PMCID: PMC2804953          DOI: 10.1109/42.712122

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  26 in total

1.  Changes in quantitative SPECT thallium-201 results associated with the use of energy-weighted acquisition.

Authors:  J L Floyd; R B Mann; A Shaw
Journal:  J Nucl Med       Date:  1991-05       Impact factor: 10.057

2.  A dual-photopeak window method for scatter correction.

Authors:  M A King; G J Hademenos; S J Glick
Journal:  J Nucl Med       Date:  1992-04       Impact factor: 10.057

3.  Analysis of the reconstructibility and noise properties of scattered photons in 99mTc SPECT.

Authors:  D J Kadrmas; E C Frey; B M Tsui
Journal:  Phys Med Biol       Date:  1997-12       Impact factor: 3.609

4.  Transmission-based scatter correction of 180 degrees myocardial single-photon emission tomographic studies.

Authors:  B F Hutton; A Osiecki; S R Meikle
Journal:  Eur J Nucl Med       Date:  1996-10

5.  Generalized scatter correction method in SPECT using point scatter distribution functions.

Authors:  P Msaki; B Axelsson; C M Dahl; S A Larsson
Journal:  J Nucl Med       Date:  1987-12       Impact factor: 10.057

6.  Deconvolution of Compton scatter in SPECT.

Authors:  C E Floyd; R J Jaszczak; K L Greer; R E Coleman
Journal:  J Nucl Med       Date:  1985-04       Impact factor: 10.057

7.  Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT.

Authors:  T Ichihara; K Ogawa; N Motomura; A Kubo; S Hashimoto
Journal:  J Nucl Med       Date:  1993-12       Impact factor: 10.057

8.  A transmission-dependent method for scatter correction in SPECT.

Authors:  S R Meikle; B F Hutton; D L Bailey
Journal:  J Nucl Med       Date:  1994-02       Impact factor: 10.057

9.  Improved SPECT quantification using compensation for scattered photons.

Authors:  R J Jaszczak; K L Greer; C E Floyd; C C Harris; R E Coleman
Journal:  J Nucl Med       Date:  1984-08       Impact factor: 10.057

10.  EM reconstruction algorithms for emission and transmission tomography.

Authors:  K Lange; R Carson
Journal:  J Comput Assist Tomogr       Date:  1984-04       Impact factor: 1.826

View more
  6 in total

1.  The value and practice of attenuation correction for myocardial perfusion SPECT imaging: a joint position statement from the American Society of Nuclear Cardiology and the Society of Nuclear Medicine.

Authors:  Robert C Hendel; James R Corbett; S James Cullom; E Gordon DePuey; Ernest V Garcia; Timothy M Bateman
Journal:  J Nucl Cardiol       Date:  2002 Jan-Feb       Impact factor: 5.952

2.  Rapid Optimization of SPECT Scatter Correction Using Model LROC Observers.

Authors:  Santosh Kulkarni; Parmeshwar Khurd; Lili Zhou; Gene Gindi
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007

3.  Development of a practical image-based scatter correction method for brain perfusion SPECT: comparison with the TEW method.

Authors:  Miho Shidahara; Hiroshi Watabe; Kyeong Min Kim; Takashi Kato; Shoji Kawatsu; Rikio Kato; Kumiko Yoshimura; Hidehiro Iida; Kengo Ito
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-05-28       Impact factor: 9.236

4.  Simultaneous technetium-99m/thallium-201 SPECT imaging with model-based compensation for cross-contaminating effects.

Authors:  D J Kadrmas; E C Frey; B M Tsui
Journal:  Phys Med Biol       Date:  1999-07       Impact factor: 3.609

5.  An investigation of the trade-off between the count level and image quality in myocardial perfusion SPECT using simulated images: the effects of statistical noise and object variability on defect detectability.

Authors:  Xin He; Jonathan M Links; Eric C Frey
Journal:  Phys Med Biol       Date:  2010-08-06       Impact factor: 3.609

6.  Small field-of-view dedicated cardiac SPECT systems: impact of projection truncation.

Authors:  Jianbin Xiao; Fred J Verzijlbergen; Max A Viergever; Freek J Beekman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-09-01       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.