Relationship between contact time measurements and PTV values when using the Periotest to measure implant stability.

N Meredith1, B Friberg, L Sennerby, C Aparicio.   

Abstract

PURPOSE: The Periotest is an electronic instrument that has been advocated for the measurement of implant stability and osseointegration. The aims of this investigation were to establish the relationship between contact time and PTV values when the Periotest was used to assess implants in vivo and in vitro, and to investigate the influence of the striking height of the Periotest handpiece and the length of implant abutment.
MATERIALS AND METHODS: The accelerometer signal from a Periotest was captured and compared with the resulting PTV value. In vitro measurements of PTV and contract time were performed on a 3-mm abutment that had been attached to a 15-mm implant luted into an aluminium block, and were repeated on a patient in vivo. further measurements were made of the abutments of six implants in turn in the maxilla of the same patient. The standard abutment lengths on the implants were 3, 4 (x 2), 5.5 (x 2), and 7 mm.
RESULTS: The results indicated that there was a linear relationship between contact time and PTV value for implants measured in vitro and in vivo. Greater scatter of the in vivo data was attributed to test and patient variables including striking position, distance, and damping as a result of the presence of soft tissue. There was a linear relationship between the PTV value and the striking height for implant measurements in vivo and in vitro.
CONCLUSION: It can be concluded that the sensitivity of the Periotest to clinical variables including striking heights and handpiece angulation limit the application of the instrument as a clinical diagnostic aid to measure implant stability.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9728122

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  12 in total

1.  The influence of micro and macro-geometry in term of bone-implant interface in two implant systems: an histomorphometrical study.

Authors:  A Rocci; R Calcaterra; M DI Girolamo; M Rocci; C Rocci; L Baggi
Journal:  Oral Implantol (Rome)       Date:  2016-07-23

2.  A stepwise under-prepared osteotomy technique improves primary stability in shallow-placed implants: a preliminary study for simultaneous vertical ridge augmentation.

Authors:  Daisuke Ueno; Kei Nakamura; Kousuke Kojima; Takeshi Toyoshima; Hideaki Tanaka; Kazuhiko Ueda; Kiyoshi Koyano; Toshiro Kodama
Journal:  Odontology       Date:  2017-08-02       Impact factor: 2.634

3.  Immediate Versus Delayed Loading of Implant for Replacement of Missing Mandibular First Molar: A Randomized Prospective Six Years Clinical Study.

Authors:  Prudhvi Raj Lakshmi Venkata Chidagam; Vijaya Chandra Gande; Sravanthi Yadlapalli; Ramani Yarlagadda Venkata; Sudheer Kondaka; Sravya Chedalawada
Journal:  J Clin Diagn Res       Date:  2017-04-01

4.  Ultrasonic Evaluation of the Bone-Implant Interface.

Authors:  Yoann Hériveaux; Vu-Hieu Nguyen; Guillaume Haïat
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

5.  Comparion of stability in titanium implants with different surface topographies in dogs.

Authors:  Nam-Sook Kim; Mong-Sook Vang; Hong-So Yang; Sang-Won Park; Ha-Ok Park; Hyun-Pil Lim
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

6.  Monitoring the Changes of Material Properties at Bone-Implant Interface during the Healing Process In Vivo: A Viscoelastic Investigation.

Authors:  Hsiang-Ho Chen; Wei-Yi Lai; Tze-Jian Chee; Ya-Hui Chan; Sheng-Wei Feng
Journal:  Biomed Res Int       Date:  2017-03-08       Impact factor: 3.411

7.  Peri-Implant Bone Behavior after Single Drill versus Multiple Sequence for Osteotomy Drill.

Authors:  Sergio Alexandre Gehrke; Raphaél Bettach; Jaime Sardá Aramburú Júnior; Juan Carlos Prados-Frutos; Massimo Del Fabbro; Jamil Awad Shibli
Journal:  Biomed Res Int       Date:  2018-04-11       Impact factor: 3.411

8.  A clinical study of alveolar bone quality using the fractal dimension and the implant stability quotient.

Authors:  Dae-Hyun Lee; Young Ku; In-Chul Rhyu; Jeong-Ug Hong; Cheol-Woo Lee; Min-Suk Heo; Kyung-Hoe Huh
Journal:  J Periodontal Implant Sci       Date:  2010-02-28       Impact factor: 2.614

9.  An in vitro evaluation of resonant frequency analysis to measure fixed bridge stability.

Authors:  Khaled Omer; Kathryn Fox; Deniel Palermo; Liam Boyle; Callum Youngson
Journal:  BDJ Open       Date:  2015-10-23

10.  Comparison of Resonance Frequency Analysis and of Quantitative Ultrasound to Assess Dental Implant Osseointegration.

Authors:  Romain Vayron; Vu-Hieu Nguyen; Benoît Lecuelle; Hugues Albini Lomami; Jean-Paul Meningaud; Romain Bosc; Guillaume Haiat
Journal:  Sensors (Basel)       Date:  2018-05-02       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.