| Literature DB >> 9727946 |
H M Gach1, I J Lowe, D P Madio, A Caprihan, S A Altobelli, D O Kuethe, E Fukushima.
Abstract
MRI systems often use magnetic field gradient and shim pulse-shaping networks (pre-emphasis) to correct for magnetic field distortions caused by eddy currents. A pre-emphasis system that uses up to 16 fixed resistor-capacitor (RC) time constants per channel with programmable amplitude coefficients is described. The magnetic fields induced by the pre-emphasis RC time constants serve as a set of basis functions for compensating eddy-current fields induced by the gradient set. The resultant time-varying magnetic field gradient accurately reflects the gradient specified by the pulse programmer. Reductions in eddy-current fields are demonstrated for actively shielded and unshielded gradient sets.Mesh:
Year: 1998 PMID: 9727946 DOI: 10.1002/mrm.1910400313
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668