Electrical proarrhythmia: induction of inappropriate atrial therapies due to far-field R wave oversensing in a new dual chamber defibrillator.

C Wolpert1, W Jung, C Scholl, S Spehl, J Cyran, B Lüderitz.   

Abstract

This case report describes delivery of atrial therapies during a sinus tachycardia in a new dual chamber implantable cardioverter defibrillator inappropriately caused by far-field oversensing of ventricular beats in the atrial channel. Upon classification of the PR interval pattern, the rate criterion for an atrial tachycardia was fulfilled, and the device initiated high-frequency burst pacing as the first stage of programmed tiered atrial therapies. Atrial fibrillation subsequently was induced by high-frequency burst pacing, and eventually a programmed 10-J shock was delivered for successful termination of atrial fibrillation. The phenomenon of far-field oversensing of ventricular beats could be repeatedly observed during exercise testing and abolished by decreasing the atrial sensitivity.

Entities:  

Mesh:

Year:  1998        PMID: 9727665     DOI: 10.1111/j.1540-8167.1998.tb00126.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  8 in total

1.  Inappropriate implantable defibrillator discharge caused by a retained pacemaker lead fragment.

Authors:  L Lickfett; C Wolpert; W Jung; S Spehl; L Pizzulli; B Esmailzadeh; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

2.  Atrial defibrillation at the millennium: new challenges for evolving technology.

Authors:  R Boccadamo
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

Review 3.  Implantable dual-chamber cardioverter-defibrillator-pacemaker.

Authors:  D Pfeiffer; M Mende; A Hagendorff
Journal:  Curr Cardiol Rep       Date:  2000-07       Impact factor: 2.931

4.  Effect of different location of atrial lead position on nearfield and farfield electrograms in dual chamber pacemaker-defibrillators.

Authors:  B K Kantharia; S L Wilbur; F A Padder; J C Pennington ; F L Samuels; C M Movsowitz; S E Hessen; C Saari; S P Kutalek
Journal:  J Interv Card Electrophysiol       Date:  2001-03       Impact factor: 1.900

5.  Dual chamber implantable cardioverter defibrillator benefits and limitations.

Authors:  K Fan; K Lee; C P Lau
Journal:  J Interv Card Electrophysiol       Date:  1999-10       Impact factor: 1.900

6.  Implantable atrial defibrillator and detection of atrial flutter.

Authors:  M Schwartz; C Maglio; M Akhtar; J Sra
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

Review 7.  Algorithms for better arrhythmia discrimination in implantable cardioverter defibrillators.

Authors:  R Schimpf; C Wolpert; B Lüderitz
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

Review 8.  The role of pacing in rhythm control and management of atrial fibrillation.

Authors:  John Silberbauer; Neil Sulke
Journal:  J Interv Card Electrophysiol       Date:  2007-05-02       Impact factor: 1.759

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.