Literature DB >> 9726311

Hybrid total hip arthroplasty in patients under the age of fifty: a five- to ten-year follow-up.

S M Sporer1, J J Callaghan, J P Olejniczak, D D Goetz, R C Johnston.   

Abstract

In 37 patients, 45 total hip replacements were performed using contemporary cementing techniques, an uncemented Harris-Galante I acetabular component and a cemented precoated Iowa femoral component in patients under the age of 50 at the time of their surgery; 36 patients with 43 hybrid hips were living, 1 patient with 2 hybrid hips was decreased. No patients were lost to follow-up. At 5- to 10-year follow-up, eight hips were revised for aseptic loosening. No acetabular components, and eight femoral components (18%) were revised for aseptic loosening. When looking at radiographic results, including revision as well as those components that were probably or definitely loose on radiographs, 0 acetabular components and 11 femoral components (24%) were radiographically loose. These results demonstrate the excellent durability of the uncemented Harris-Galante acetabular component in the younger patient. However, the Iowa grit-blasted methyl methacrylate precoated femoral component had a magnitude increase in the prevalence of revision for aseptic femoral loosening when compared to the senior author's long-term Charnley results in this age group. The authors attribute the failure to the rough surface finish applied to the femoral component. However, the polymethyl methacrylate proximal precoating and the femoral component design may also contribute to the femoral failures.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9726311     DOI: 10.1016/s0883-5403(98)90045-6

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  6 in total

1.  Bousquet dual mobility socket for patient under fifty years old. More than twenty year follow-up of one hundred and thirty one hips.

Authors:  Remi Philippot; Thomas Neri; Bertrand Boyer; Brice Viard; Frederic Farizon
Journal:  Int Orthop       Date:  2017-01-16       Impact factor: 3.075

2.  Favourable mid-term results of the VerSys CT polished cemented femoral stem for total hip arthroplasty.

Authors:  Alejandro González Della Valle; Fernando Comba; Adriana Zoppi; Eduardo A Salvati
Journal:  Int Orthop       Date:  2006-03-31       Impact factor: 3.075

Review 3.  Hip resurfacing versus total hip arthroplasty: a systematic review comparing standardized outcomes.

Authors:  Deborah A Marshall; Karen Pykerman; Jason Werle; Diane Lorenzetti; Tracy Wasylak; Tom Noseworthy; Donald A Dick; Greg O'Connor; Aish Sundaram; Sanne Heintzbergen; Cy Frank
Journal:  Clin Orthop Relat Res       Date:  2014-04-04       Impact factor: 4.176

4.  Direct comparison of polyethylene wear in cemented and uncemented acetabular cups.

Authors:  Håvard Bjerkholt; Oystein Høvik; Olav Reikerås
Journal:  J Orthop Traumatol       Date:  2010-09-02

5.  Congenital hip dysplasia treated by total hip arthroplasty using cementless tapered stem in patients younger than 50 years old: results after 12-years follow-up.

Authors:  Cesare Faldini; Maria Teresa Miscione; Mohammadreza Chehrassan; Francesco Acri; Camilla Pungetti; Michele d'Amato; Deianira Luciani; Sandro Giannini
Journal:  J Orthop Traumatol       Date:  2011-11-24

6.  19 years outcome after cementless total hip arthroplasty with spongy metal structured implants in patients younger than 65 years.

Authors:  Ludger Gerdesmeyer; Munjed Al Muderis; Hans Gollwitzer; Norbert Harrasser; Martin Stukenberg; Maria-Angela Clifford; Andreas Toepfer
Journal:  BMC Musculoskelet Disord       Date:  2016-10-18       Impact factor: 2.362