Literature DB >> 9718530

Differences in observer variability of ultrasound measurements of the proximal and distal abdominal aorta.

H J Pleumeekers1, A W Hoes, P G Mulder, E van der Does, A Hofman, J S Laméris, D E Grobbee.   

Abstract

OBJECTIVE: To assess the observer variability of ultrasound measurements of the abdominal aorta and to study whether observer variability is influenced by the site of measurement or cardiovascular risk factors.
SETTING: Population based screening programme for abdominal aortic aneurysms.
METHODS: For 135 subjects taking part in a screening programme for abdominal aortic aneurysms, two of the three ultrasonographers measured the distal and proximal ultrasound diameter of the abdominal aorta, using B-mode ultrasound, according to the Rotterdam study scanning protocol.
RESULTS: The mean difference between two different observers was 0.06 mm (95% CI-0.15 to 0.27) for measurements of the distal aorta and 0.32 mm (95% CI 0.09 to 0.55) for the proximal aorta. Maximal differences between observers for measurements of both the distal and proximal aortic diameter were 4.0 mm. Interobserver variability in the proximal and distal measurements of the abdominal aorta was not related to the level of the major cardiovascular risk indicators. However, interobserver variability in ultrasound measurements of the proximal aorta increased with increasing waist circumference and increasing diameter of the proximal aorta.
CONCLUSION: Ultrasonographic readings of the distal and proximal aortic measurements can be interpreted within a range of plus or minus 3 mm. Ultrasound measurements are more accurate for the distal than for the proximal measurement. Definition of the aortic diameter based on a combination of both distal and proximal measurement may be more accurate.

Entities:  

Mesh:

Year:  1998        PMID: 9718530     DOI: 10.1136/jms.5.2.104

Source DB:  PubMed          Journal:  J Med Screen        ISSN: 0969-1413            Impact factor:   2.136


  5 in total

1.  Office-based ultrasound screening for abdominal aortic aneurysm.

Authors:  Beau Blois
Journal:  Can Fam Physician       Date:  2012-03       Impact factor: 3.275

2.  Ultrasound diagnostics of the abdominal aorta: English version.

Authors:  W Schäberle; L Leyerer; W Schierling; K Pfister
Journal:  Gefasschirurgie       Date:  2015

3.  The reproducibility of measuring maximum abdominal aortic aneurysm diameter from ultrasound images.

Authors:  Evan O Matthews; Jenna Pinchbeck; Kylie Elmore; Rhondda E Jones; Joseph V Moxon; Jonathan Golledge
Journal:  Ultrasound J       Date:  2021-02-26

4.  Reproducibility of the SPI-US protocol for ultrasound diameter measurements of the Posterior Circumflex Humeral Artery and Deep Brachial Artery: an inter-rater reliability study.

Authors:  Daan van de Pol; Sena Alaeikhanehshir; P Paul F M Kuijer; Aart Terpstra; Marja J C Pannekoek-Hekman; R Nils Planken; Mario Maas
Journal:  Eur Radiol       Date:  2015-12-10       Impact factor: 5.315

5.  Abdominal aorta measurements by a handheld ultrasound device compared with a conventional cart-based ultrasound machine.

Authors:  Abdulrahman M Alfuraih; Abdulaziz I Alrashed; Saleh O Almazyad; Mohammed J Alsaadi
Journal:  Ann Saudi Med       Date:  2021-12-02       Impact factor: 1.526

  5 in total