Determination of cardiac output during mechanical ventilation by electrical bioimpedance or thermodilution in patients with acute lung injury: effects of positive end-expiratory pressure.

M Genoni1, P Pelosi, J A Romand, A Pedoto, T Moccetti, R Malacrida.   

Abstract

OBJECTIVE: To evaluate the usefulness of transthoracic electrical bioimpedance in sedated and paralyzed patients with acute lung injury during mechanical ventilation with and without early application of positive end-expiratory pressure (PEEP).
DESIGN: Prospective, repeated-measures study.
SETTING: University-affiliated intensive care center. PATIENTS: Ten patients with acute lung injury.
INTERVENTIONS: Simultaneous, three-paired cardiac output (CO) measurements by transthoracic electrical bioimpedance (TEB) and thermodilution (TD) were made at 0 and 15 cm H2O of PEEP.
MEASUREMENTS AND MAIN RESULTS: The average of the TD-CO measurements was 7.22 +/- 2.12 (SD) L/min during 0 cm H2O of positive end-expiratory pressure (ZEEP), and 6.91 +/- 1.72 L/min during PEEP (NS). The average of the TEB-CO measurements was 4.48 +/- 1.37 L/min during ZEEP, and 6.03 +/- 2.03 L/min during PEEP (p < .05). For each level of PEEP, bias and precision between methods were calculated. Bias calculations between TD-CO and TEB-CO ranged from -1.54 +/- 7.02 L/min at ZEEP to -2.52 +/- 4.28 L/ min at PEEP, and -2.47 +/- 6.09 L/min for mixed data at ZEEP and PEEP. There was no significant correlation between the percent change with PEEP in TEB-CO and TD-CO (r2 =.05, NS).
CONCLUSIONS: In patients with acute lung injury: a) the agreement between TEB-CO and TD-CO measurements is poor; b) agreement is not clinically improved by application of PEEP; and c) TEB cannot monitor trends in CO.

Entities:  

Mesh:

Year:  1998        PMID: 9710107     DOI: 10.1097/00003246-199808000-00035

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  4 in total

1.  Stroke volume equation for impedance cardiography.

Authors:  D P Bernstein; H J M Lemmens
Journal:  Med Biol Eng Comput       Date:  2005-07       Impact factor: 2.602

2.  Physiologic and Clinical Principles behind Noninvasive Resuscitation Techniques and Cardiac Output Monitoring.

Authors:  Anthony M Napoli
Journal:  Cardiol Res Pract       Date:  2011-08-16       Impact factor: 1.866

Review 3.  Equipment review: the success of early goal-directed therapy for septic shock prompts evaluation of current approaches for monitoring the adequacy of resuscitation.

Authors:  Scott R Gunn; Mitchell P Fink; Benjamin Wallace
Journal:  Crit Care       Date:  2005-05-27       Impact factor: 9.097

4.  Comparative study of cardiac output measurement by regional impedance cardiography and thermodilution method in patients undergoing off pump coronary artery bypass graft surgery.

Authors:  Amrita Guha; Dheeraj Arora; Yatin Mehta
Journal:  Ann Card Anaesth       Date:  2022 Jul-Sep
  4 in total