Literature DB >> 9686801

Bicarbon valve -- European multicenter clinical evaluation.

J B Borman1, W G Brands, L Camilleri, M Cotrufo, W Daenen, I Gandjbakhch, C Infantes, A Khayat, F Laborde, A Pellegrini, A Piwnica, B Reichart, R Sharony, R Walesby, H Warembourg.   

Abstract

OBJECTIVE: Fifteen collaborating centers in eight countries present their pooled experience with the new Bicarbon bileaflet valve.
METHODS: Between 4/90 and 4/96, 1351 patients, 806 males and 545 females, aged 10 to 83, mean 58.4 +/- 12.4, underwent valve implantation. OPERATIONS: aortic valve replacement (AVR), 726; mitral valve replacement (MVR), 475; double valve replacement (DVR), 150. Additional procedures: CABG, 211; TV repair, 64; other, 152.
RESULTS: Mortality: 67 early (seven valve related) and 56 late (40 valve related). Valve thrombosis: six obstructive, three non-obstructive; embolism: nine major cerebral, 37 other. Major bleeding: 29. Hemolysis: two clinically significant. Non-structural dysfunction: 24 paravalvular leaks, one leaflet interference. No structural failure! Endocarditis: 24. Reoperation 48: 22 non-structural dysfunctions, 14 endocarditis, seven thrombosis and embolism, five other. Estimated 5-year freedom from valve-related deaths is 97.2% for AVR and 92.4% for MVR; 4-year freedom from valve related deaths for DVR is 90.5%. Mean calculated NYHA improvement is 1.24.
CONCLUSIONS: The Bicarbon mechanical prosthesis is well designed, durable, has good hemodynamic features and an acceptably low incidence of complications.

Entities:  

Mesh:

Year:  1998        PMID: 9686801     DOI: 10.1016/s1010-7940(98)00074-8

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  4 in total

1.  Clinical experience with the Bicarbon heart valve prosthesis.

Authors:  Yoshio Misawa; Tsutomu Saito; Hiroaki Konishi; Shin-Ichi Oki; Yuichiro Kaminishi; Yasuhito Sakano; Hideki Morita; Kei Aizawa
Journal:  J Cardiothorac Surg       Date:  2007-01-25       Impact factor: 1.637

2.  Long-term experience with the Sorin Bicarbon and Edwards Mira mechanical valve prostheses in the mitral position.

Authors:  Toshihiro Fukui; Toshihiko Shibata; Yasuyuki Sasaki; Hidekazu Hirai; Manabu Motoki; Yosuke Takahashi; Shigefumi Suehiro
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-03

3.  Fifteen-year experience with the Bicarbon heart valve prosthesis in a single center.

Authors:  Yoshio Misawa; Arata Muraoka; Shin-ichi Ohki; Kei Aizawa; Koji Kawahito; Tsutomu Saito; Hirotaka Sato; Ippei Takazawa; Soki Kurumisawa; Hirohiko Akutsu; Akira Sugaya
Journal:  J Cardiothorac Surg       Date:  2015-06-28       Impact factor: 1.637

4.  Nineteen Years of Single Institute Experiences with Sorin Bicarbon Prosthesis.

Authors:  Hung Dung Van
Journal:  Ann Thorac Cardiovasc Surg       Date:  2019-03-12       Impact factor: 1.520

  4 in total