Literature DB >> 9683944

Measurement of axial length in the calculation of intraocular lens power.

P S Raj1, B Ilango, A Watson.   

Abstract

PURPOSE: Accurate measurement of ocular axial length is essential for accurate intraocular lens (IOL) power calculation. Although it is common practice to average several axial length measurements to improve accuracy, it has been suggested that a single high-quality A-scan ultrasonographic measurement is adequately accurate owing to the high test-retest reliability of A-scan biometry. The aim of this study was to compare the accuracy of a single high-quality A-scan measurement with that of the average of three acceptable measurements in the calculation of IOL power.
METHOD: We studied 103 eyes of 103 patients who underwent cataract-IOL surgery. All these patients underwent pre-operative ocular biometry, a standardised extracapsular cataract extraction with posterior chamber IOL implantation, and clinical refraction between 10 and 14 weeks post-operatively.
RESULTS: There was no statistically significant difference between the two study groups in measured axial length (p > 0.01), calculated emmetropic IOL power (p > 0.05) or the prediction of post-operative refraction (p > 0.99).
CONCLUSION: The use of a single high-quality axial length measurement was as accurate as the mean of three acceptable axial length measurements in the calculation of IOL power.

Entities:  

Mesh:

Year:  1998        PMID: 9683944     DOI: 10.1038/eye.1998.54

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  6 in total

1.  Refractive outcomes comparison between the Lenstar LS 900® optical biometry and immersion A-scan ultrasound.

Authors:  Palanyraj Naicker; Siva Sundralingam; Mohammadreza Peyman; Azida Juana; Nor Fadhilah Mohamad; Maung Maung Win; Angela Loo; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-07-15       Impact factor: 2.031

2.  A new non-contact optical device for ocular biometry.

Authors:  J Santodomingo-Rubido; E A H Mallen; B Gilmartin; J S Wolffsohn
Journal:  Br J Ophthalmol       Date:  2002-04       Impact factor: 4.638

3.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

4.  Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography.

Authors:  Wenwen Wang; Yaxin Miao; Giacomo Savini; Colm McAlinden; Hao Chen; Qingjie Hu; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-10-23       Impact factor: 4.379

5.  Imbalance Between Oxidative Stress and Growth Factors in Human High Myopia.

Authors:  Salvador Mérida; Vincent M Villar; Amparo Navea; Carmen Desco; María Sancho-Tello; Cristina Peris; Francisco Bosch-Morell
Journal:  Front Physiol       Date:  2020-05-14       Impact factor: 4.566

6.  Refractive Results Using a New Optical Biometry Device: Comparison With Ultrasound Biometry Data.

Authors:  Serdar Aktas; Hatice Aktas; Mehmet Tetikoglu; Hac Murat Sagdk; Fatih Özcura
Journal:  Medicine (Baltimore)       Date:  2015-12       Impact factor: 1.817

  6 in total