Determination of functional residual capacity (FRC) by multibreath nitrogen washout in a lung model and in mechanically ventilated patients. Accuracy depends on continuous dynamic compensation for changes of gas sampling delay time.

H Wrigge1, M Sydow, J Zinserling, P Neumann, J Hinz, H Burchardi.   

Abstract

OBJECTIVE: Validation of an open-circuit multibreath nitrogen washout technique (MBNW) for measurement of functional residual capacity (FRC). The accuracy of FRC measurement with and without continuous viscosity correction of mass spectrometer delay time (TD) relative to gas flow signal and the influence of baseline FIO2 was investigated.
DESIGN: Laboratory study and measurements in mechanically ventilated patients.
SETTING: Experimental laboratory and anesthesiological intensive care unit of a university hospital. PATIENTS: 16 postoperative patients with normal pulmonary function (NORM), 8 patients with acute lung injury (ALI) and 6 patients with chronic obstructive pulmonary disease (COPD) were included.
INTERVENTIONS: Change of FIO2 from baseline to 1.0.
MEASUREMENTS AND MAIN RESULTS: FRC was determined by MBNW using continuous viscosity correction of TD(TDdyn), a constant TD based on the viscosity of a calibration gas mixture (TD0) and a constant TD referring to the mean viscosity between onset and end of MBNW (TDmean). Using TDdyn, the mean deviation between 15 measurements of three different lung model FRCs (FRCmeasured) and absolute volumes (FRCmodel) was 0.2%. For baseline FIO2 ranging from 0.21 to 0.8, the mean deviation between FRCmeasured and FRCmodel was -0.8%. However, depending on baseline FIO2, the calculation of FRC using TDmean and TD0 increased the mean deviation between FRCmeasured and FRCmodel to 2-4% and 8-12%, respectively. In patients (n = 30) the average repeatability coefficient was 6.0%. FRC determinations with TDmean and TD0 were 0.8-13.3% and 4.2-23.9% (median 2.7% and 8.7%) smaller than those calculated with TDdyn.
CONCLUSION: A dynamic viscosity correction of TD improves the accuracy of FRC determinations by MBNW considerably, when gas concentrations are measured in a sidestream. If dynamic TD correction cannot be performed, the use of constant TDmean might be suitable. However, in patient measurements this can cause an FRC underestimation of up to 13%.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9660266     DOI: 10.1007/s001340050601

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  13 in total

1.  What is the "best PEEP" in chronic obstructive pulmonary disease?

Authors:  H Wrigge; C Putensen
Journal:  Intensive Care Med       Date:  2000-09       Impact factor: 17.440

2.  Measurement of lung volume in mechanically ventilated monkeys with an ultrasonic flow meter and the nitrogen washout method.

Authors:  Andreas Schibler; Jürg Hammer; Ruedi Isler; Christian Buess; Christopher J L Newth
Journal:  Intensive Care Med       Date:  2003-10-07       Impact factor: 17.440

3.  Relating indices of inert gas washout to localised bronchoconstriction.

Authors:  Jennine H Mitchell; Eric A Hoffman; Merryn H Tawhai
Journal:  Respir Physiol Neurobiol       Date:  2012-07-05       Impact factor: 1.931

Review 4.  [Recruitment maneuvers for patients with lung failure. When, how, whether or not?].

Authors:  J Hinz; O Moerer; M Quintel
Journal:  Anaesthesist       Date:  2005-11       Impact factor: 1.041

5.  Slow moderate pressure recruitment maneuver minimizes negative circulatory and lung mechanic side effects: evaluation of recruitment maneuvers using electric impedance tomography.

Authors:  Helena Odenstedt; Sophie Lindgren; Cecilia Olegård; Karin Erlandsson; Sven Lethvall; Anders Aneman; Ola Stenqvist; Stefan Lundin
Journal:  Intensive Care Med       Date:  2005-09-22       Impact factor: 17.440

6.  Mechanical ventilator for delivery of ¹⁷O₂ in brief pulses.

Authors:  J E Baumgardner; E A Mellon; D R Tailor; K Mallikarjunarao; A Borthakur; R Reddy
Journal:  Open Biomed Eng J       Date:  2008-09-24

7.  Validation and clinical application of a first order step response equation for nitrogen clearance during FRC measurement.

Authors:  Gary Choncholas; Soren Sondergaard; Erkki Heinonen
Journal:  J Clin Monit Comput       Date:  2007-11-15       Impact factor: 2.502

8.  Lung volume in mechanically ventilated patients: measurement by simplified helium dilution compared to quantitative CT scan.

Authors:  Nicolò Patroniti; Giacomo Bellani; Annamaria Manfio; Elena Maggioni; Angela Giuffrida; Giuseppe Foti; Antonio Pesenti
Journal:  Intensive Care Med       Date:  2004-01-09       Impact factor: 17.440

9.  Measurement of functional residual capacity by helium dilution during partial support ventilation: in vitro accuracy and in vivo precision of the method.

Authors:  Fabiano Di Marco; Lidia Rota Sperti; Barbara Milan; Riccardo Stucchi; Stefano Centanni; Laurent Brochard; Roberto Fumagalli
Journal:  Intensive Care Med       Date:  2007-08-17       Impact factor: 17.440

10.  Determination of functional residual capacity by oxygen washin-washout: a validation study.

Authors:  Stefan Maisch; Stephan H Boehm; Dieter Weismann; Hajo Reissmann; Marcus Beckmann; Bernd Fuellekrug; Andreas Meyer; Jochen Schulte Am Esch
Journal:  Intensive Care Med       Date:  2007-03-15       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.